首页> 外文期刊>Journal of Medical Imaging and Health Informatics >Design and Testing of a Multi-Channel H-1/He-3 Double-Nuclear Transmit Coil Array for Magnetic Resonance Imaging at an Ultra-Low Field
【24h】

Design and Testing of a Multi-Channel H-1/He-3 Double-Nuclear Transmit Coil Array for Magnetic Resonance Imaging at an Ultra-Low Field

机译:用于超低场的多通道H-1 / HE-3双核传动线圈阵列的设计和测试

获取原文
获取原文并翻译 | 示例
           

摘要

The application of polarized noble gases in lung functional magnetic resonance imaging (fMRI) has garnered attention due its capacity to produce a high-resolution image with no radiation. For image signal acquisition, the transmit coil requires a higher performance whereas the traditional single frequency transmit coil is unsuitable. This paper presents a novel radio frequency (RF) coil for application in a H-1/He-3 MRI system at the ultra-low field of 0.06 T. For the more complex model of a 4-channel double-nuclear H-1/He-3 coil, traditional electromagnetic simulation is not applicable due to its long computing time. The electromagnetic (EM) field and RF circuit co-simulation method were implemented to accelerate the analysis, resulting in an effective evaluation. The co-simulation and test results showed that the proposed model has the potential for imaging of the lung using the H-1/He-3 MRI system at an ultra-low field.
机译:偏振惰性气体在肺功能磁共振成像(FMRI)中的应用由于其产能为不具有辐射的高分辨率图像而受到关注。 对于图像信号采集,发射线圈需要更高的性能,而传统的单频透射线圈是不合适的。 本文提出了一种新型射频(RF)线圈,用于在0.06T的超低场的H-1 / HE-3 MRI系统中应用,用于更复杂的4通道双核H-1模型 / HE-3线圈,传统电磁仿真由于其长期计算时间而不适用。 实施电磁(EM)场和RF电路共模拟方法以加速分析,导致有效的评估。 共模和测试结果表明,所提出的模型在超低场中使用H-1 / HE-3 MRI系统对肺部成像的可能性。

著录项

  • 来源
  • 作者单位

    Hefei Univ Technol Dept Elect &

    Automat Engn Hefei 230000 Anhui Peoples R China;

    Chinese Acad Sci Shenzhen Inst Adv Technol Shenzhen 518055 Peoples R China;

    Hefei Univ Technol Dept Elect &

    Automat Engn Hefei 230000 Anhui Peoples R China;

    Chinese Acad Sci Shenzhen Inst Adv Technol Shenzhen 518055 Peoples R China;

    Hefei Univ Technol Dept Elect &

    Automat Engn Hefei 230000 Anhui Peoples R China;

    Chinese Acad Sci Shenzhen Inst Adv Technol Shenzhen 518055 Peoples R China;

    Chinese Acad Sci Shenzhen Inst Adv Technol Shenzhen 518055 Peoples R China;

    Chinese Acad Sci Shenzhen Inst Adv Technol Shenzhen 518055 Peoples R China;

    Univ Calif San Francisco Dept Radiol &

    Biomed Imaging San Francisco CA 94158 USA;

    Chinese Acad Sci Shenzhen Inst Adv Technol Shenzhen 518055 Peoples R China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 放射卫生;
  • 关键词

    MRI; Noble Gases; Double-Nuclear; Transmit Coil Array; EM Field and RF Circuit Co-Simulation;

    机译:MRI;高贵气体;双核;传输线圈阵列;EM场和RF电路共模;