Clinical Results and Mechanical Properties of the Carotid CGUARD Double-Layered Embolic Prevention Stent

机译:颈动脉颈双层栓塞预防支架的临床结果与力学性能

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To report early clinical outcomes with a novel double-layer stent for the internal carotid artery (ICA) and the in vitro investigation of the stent's mechanical properties. Methods: A prospective single-center study enrolled 30 consecutive patients (mean age 73.1 +/- 6.3 years; 21 men) with symptomatic (n=25) or high-grade (n=5) ICA stenosis treated with the new double-layer carotid CGUARD Embolic Prevention System (EPS) stent, which has an inner open-cell nitinol design with an outer closed-cell polyethylene terephthalate layer. The average stenosis of the treated arteries was 84.1%+/- 7.9% with a mean lesion length of 16.6 +/- 2.1 mm. In the laboratory, 8x40-mm stents where tested in vitro with respect to their radial force during expansion, the bending stiffness of the stent system and the expanded stent, as well as the collapse pressure in a thin and flexible sheath. The wall adaptation was assessed using fluoroscopy after stent release in step and curved vessel models. Results: The stent was successfully implanted in all patients. No peri- or postprocedural complications occurred; no minor or major stroke was observed in the 6-month follow-up. The bending stiffness of the expanded stent was 63.1 Nmm(2) and (not unexpectedly) was clearly lower than that of the stent system (601.5 Nmm(2)). The normalized radial force during expansion of the stent to 7.0 mm, consistent with in vivo sizing, was relatively high (0.056 N/mm), which correlates well with the collapse pressure of 0.17 bars. Vessel wall adaptation was harmonic and caused no straightening of the vessel after clinical application. Conclusion: Because of its structure, the novel CGUARD EPS stent is characterized by a high flexibility combined with a high radial force and very good plaque coverage. These first clinical results demonstrate a very safe implantation behavior without any stroke up to 6 months after the procedure.
机译:目的:以内部颈动脉(ICA)的新型双层支架报告早期临床结果以及支架的机械性能的体外调查。方法:前瞻性单中心研究连续30名患者(平均年龄73.1 +/- 6.3岁; 21人),用新的双层处理的症状(n = 25)或高档(n = 5)ICA狭窄颈动脉爪栓塞预防系统(EPS)支架,其具有内部开孔镍钛合金,具有外闭细胞聚对苯二甲酸乙二醇酯层。经处理的动脉的平均狭窄为84.1%+ / - 7.9%,平均病变长度为16.6 +/- 2.1 mm。在实验室,8×40mm支架,其中在膨胀期间相对于其径向力进行体外测试,支架系统的弯曲刚度和膨胀支架,以及薄且柔性护套中的塌陷压力。在步骤和弯曲容器模型中支架释放后使用透视评估壁适应。结果:支架在所有患者中成功植入。没有发生Peri或后期并发症;在6个月的随访中没有观察到未成年或重大卒中。膨胀支架的弯曲刚度为63.1nmm(2),(不出意外)明显低于支架系统(601.5 nmm(2))。支架的膨胀期间至7.0mm的归一化径向力相对较高(0.056N / mm),其与0.17巴的塌陷压力很好。血管壁适应伴谐波并导致临床应用后的血管伸直。结论:由于其结构,新型CGUARD EPS支架的特点是具有高径向力和非常好的斑块覆盖率的高柔韧性。这些第一个临床结果表明,在手术后6个月内没有任何中风,展示了非常安全的植入行为。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号