The potential benefits of preserving high-frequency spectral cues created by the pinna in hearing-aid fittings were investigated in a combined laboratory and field test. In a single-blind crossover design, two settings of an experimental hearing aid were compared. One setting was characterized by a pinna cue-preserving microphone position, whereas the other was characterized by a microphone position not preserving pinna cues. Participants were allowed I month of acclimatization to each setting before measurements of localization and spatial release from speech-on-speech masking were completed in the laboratory. Real-world experience with the two settings was assessed by means of questionnaires. Seventeen participants with mild to moderate sensorineural hearing impairments completed the study. An inconsistent pinna cue benefit pattern was observed across the outcome measures. In the localization test, the pinna cue-preserving setting provided a significant mean reduction of 22° in the root mean square (RMS) error in the front-back dimension, with 13 of the 17 participants showing a reduction of at least 15°. No significant mean difference in RMS error between settings was observed in the left-right dimension. No significant differences between settings were observed in the spatial-unmasking test conditions. The questionnaire data indicated a small, but nonsignificant, benefit of the pinna cue-preserving setting in certain real-life situations, which corresponded with a general preference for that setting. No significant real-life localization benefit was observed. The results suggest that preserving pinna cues can offer benefit in some conditions for individual hearing-aid users with mild to moderate hearing loss and is unlikely to harm performances for the rest.
机译:在一个组合的实验室和现场试验中,研究了在助剂配件中保留由PinNA产生的高频光谱线索的潜在益处。在单盲交叉设计中,比较了两种实验助听器的设置。一个设置的特征在于PinNA提示保存麦克风位置,而另一个设置的特征在于不保留磁质素线索的麦克风位置。在实地化和语音上屏蔽的定位和空间释放的测量之前,将参与者允许参与者适应每个环境。通过问卷评估两个设置的真实世界经验。 17名与温和至中等感觉障碍的参与者完成了这项研究。在结果措施中观察到不一致的Pinna Cue益处模式。在本地化测试中,脊线管保存设置在前后尺寸中的根均线(RMS)误差中的22°的显着平均值减小,17名参与者中有13个,表示降低至少15°。在左右尺寸中观察到设置之间的RMS误差的显着平均差异。在空间 - 揭露试验条件下观察到设置之间没有显着差异。调查问卷数据表明,在某些真实情况下,脊柱管保存设置的小而非显着,涉及与该设置的一般偏好相对应的。没有观察到明显的真实定位福利。结果表明,保留的Pinna提示可以在某些条件下为各个助听器的条件提供益处,以轻度到中度听力损失,并且不太可能危害其余的表现。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号