A validation study of multicenter diffusion tensor imaging: Reliability of fractional anisotropy and diffusivity values

机译:多中心扩散张量成像的验证研究:分数各向异性和扩散率值的可靠性

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND PURPOSE: DTI is increasingly being used as a measure to study tissue damage in several neurologic diseases. Our aim was to investigate the comparability of DTI measures between different MR imaging magnets and platforms. MATERIALS AND METHODS: Two healthy volunteers underwent DTI on five 3T MR imaging scanners (3 Trios and 2 Signas) by using a matched 33 noncollinear diffusion-direction pulse sequence. Within each subject, a total of 16 white matter (corpus callosum, periventricular, and deep white matter) and gray matter (cortical and deep gray) ROIs were drawn on a single image set and then were coregistered to the other images. Mean FA, ADC, and longitudinal and transverse diffusivities were calculated within each ROI. Concordance correlations were derived by comparing ROI DTI values among each of the 5 magnets. RESULTS: Mean concordance for FA was 0.96; for both longitudinal and transverse diffusivities, it was 0.93; and for ADC, it was 0.88. Mean scan-rescan concordance was 0.96-0.97 for all DTI measures. Concordance correlations within platforms were, in general, better than those between platforms for all DTI measures (mean concordance of 0.96). CONCLUSIONS: We found that a 3T magnet and high-angular-resolution pulse sequence yielded comparable DTI measurements across different MR imaging magnets and platforms. Our results indicate that FA is the most comparable measure across magnets, followed by individual diffusivities. The comparability of DTI measures between different magnets supports the feasibility of multicentered clinical trials by using DTI as an outcome measure.
机译:背景与目的:DTI被越来越多地用作研究几种神经系统疾病中组织损伤的一种手段。我们的目的是研究不同MR成像磁体和平台之间DTI措施的可比性。材料与方法:两名健康志愿者使用匹配的33个非共线扩散方向脉冲序列在五台3T MR成像扫描仪(3个Trios和2个Signas)上进行了DTI。在每个对象中,在单个图像集上总共绘制了16种白质(call体,脑室周围和深部白质)和灰质(皮质和深灰)ROI,然后将其与其他图像共同配准。在每个ROI内计算平均FA,ADC以及纵向和横向扩散率。通过比较5个磁体中每个磁体的ROI DTI值,得出一致性相关性。结果:FA的平均一致性为0.96;纵向和横向扩散率均为0.93;对于ADC,则为0.88。所有DTI量度的平均扫描-再扫描一致性为0.96-0.97。通常,对于所有DTI度量,平台内的一致性相关性均好于平台之间的一致性(平均一致性为0.96)。结论:我们发现3T磁体和高角度分辨率脉冲序列在不同的MR成像磁体和平台上产生了可比的DTI测量。我们的结果表明,FA是跨磁体最可比的量度,其次是单个扩散率。 DTI措施在不同磁体之间的可比性通过将DTI用作结果指标,支持了多中心临床试验的可行性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号