Performance Characteristics of the Digital Biograph Vision PET/CT System

机译:数字传销视觉宠物/ CT系统的性能特征

获取原文
获取原文并翻译 | 示例
           

摘要

This study evaluated the performance of the Biograph Vision digital PET/CT system according to the NEMA NU 2-2012 standard (published by the National Electrical Manufacturers Association [ NEMA]) to allow for a reliable, reproducible, and intersystem-comparable performance measurement. Methods: The new digital PET/CT system features silicon photomultiplier-based detectors with 3.2-mm lutetium oxyorthosilicate crystals and full coverage of the scintillator area. The PET components incorporate 8 rings of 38 detector blocks, and each block contains 4 x 2 mini blocks. Each mini block consists of a 5 x 5 lutetium oxyorthosilicate array of 3.2 x 3.2 x 20 mm crystals coupled to a silicon photomultiplier array of 16 x 16 mm, resulting in an axial field of view of 26.1 cm. In this study, PET/CT system performance was evaluated for conformation with the NEMA NU 2-2012 standard, with additional measurements described in the new NEMA NU 2-2018 standard. Spatial resolution, sensitivity, count-rate performance, accuracy of attenuation and scatter correction, image quality, coregistration accuracy, and time-of-flight performance were determined. Measurements were directly compared with results from its predecessor, the Biograph mCT Flow, using existing literature. Moreover, feasibility to comply with the European Association of Nuclear Medicine Research Ltd. (EARL) criteria was evaluated, and some illustrative patient PET images were obtained. Results: The Biograph Vision showed a transverse and axial spatial resolution of 3.6 and 3.5 mm, respectively, in full width at half maximum at a 1-cm offset from the center of the field of view (measured with a Na-22 0.25-mm point source), a NEMA sensitivity of 16.4 kcps/MBq, and a NEMA peak noise-equivalent count-rate of 306 kcps at 32 kBq/mL. Time-of-flight resolution varied from 210 to 215 as count-rate increased up to the peak noise-equivalent count-rate. The overall image contrast seen with the NEMA image quality phantom ranged from 77.2% to 89.8%. Furthermore, the system was able to comply with the current and future EARL performance criteria. Conclusion: The Biograph Vision outperforms the analog Biograph mCT Flow, and the system is able to meet European harmonizing performance standards.
机译:本研究评估了根据NEMA NU 2-2012标准的传销视觉数字PET / CT系统的性能(由国家电气制造商协会[NEMA]发布),以允许可靠,可重复和间接的性能测量。方法:新型数字PET / CT系统具有硅光电倍增管的探测器,具有3.2mm烯烯型硅酸盐晶体和闪烁体区域的完全覆盖。 PET组件包含8个探测器块的8个环,每个块包含4×2迷你块。每个迷你块由3.2×3.2×20mm晶体的5×5含氧化锂阵列组成,耦合到16×16mm的硅光电倍数阵列,导致轴向视场为26.1cm。在该研究中,评估PET / CT系统性能以符合NEMA NU 2-2012标准,并在新的NEMA NU 2-2018标准中描述了额外的测量。确定了空间分辨率,灵敏度,计数率性能,衰减和散射校正的准确性,图像质量,重铸精度和飞行时间性能。使用现有文献,与其前任的传记MCT流程的结果直接测量。此外,评估欧洲核医学研究有限公司(伯爵)标准的可行性评估,并获得了一些说明性患者PET图像。结果:传记视觉显示横向和轴向空间分辨率分别为3.6和3.5毫米,在从视野的中心的1cm偏移下的半最大宽度下为3.6和3.5毫米(用NA-22 0.25mm测量点来源),NEMA敏感性为16.4 kCPS / MBQ,NEMA峰值噪声 - 当量计数为32kBQ / mL的306kCPS。由于计数增加到峰值噪声等效计率,飞行时间分辨率从210到215变化。随着NEMA图像质量幻影的整体图像对比度范围为77.2%至89.8%。此外,该系统能够遵守当前和未来的伯爵性能标准。结论:传记视觉优于模拟传记MCT流,系统能够满足欧洲协调性能标准。

著录项

  • 来源
    《The Journal of Nuclear Medicine》 |2019年第7期|共6页
  • 作者单位

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

    Univ Groningen Univ Med Ctr Groningen Dept Nucl Med &

    Mol Imaging Groningen Netherlands;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类 放射医学;
  • 关键词

    digital detectors; PET/CT; NEMA; performance evaluation;

    机译:数字探测器;PET / CT;NEMA;绩效评估;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号