首页> 外文期刊>Sleep and biological rhythms. >Inter-scorer reliability of sleep assessment using EEG and EOG recording system in comparison to polysomnography
【24h】

Inter-scorer reliability of sleep assessment using EEG and EOG recording system in comparison to polysomnography

机译:使用EEG和EOG记录系统与多核识别相比,睡眠评估的Inter-Inter-anse评估可靠性

获取原文
获取原文并翻译 | 示例
           

摘要

There are increasing needs for self-applicable methods assessing sleep in clinical and non-clinical settings. This study aimed to investigate the inter-scorer reliability for the sleep stage scoring and for the sleep variable assessments in the portable electroencephalography (EEG) and electro-oculography (EOG) recording system. Sleep recordings were performed simultaneously with polysomnography (PSG) and portable EEG/EOG recordings in 14 healthy adults. The inter-scorer concordance of the sleep stage scoring was assessed in the two recording methods. Sleep variables were compared between the two methods using the Bland-Altman plots in each scorer. The overall inter-scorer concordance rate of sleep stage scoring for the EEG/EOG data was 85.5 %, and the Cohen's kappa was 0.80 (p < 0.001), being equivalent to those for the PSG data (89.2 %, 0.85, p < 0.001) while that for StageN1 was relatively low (EEG/ EOG 60.1 %, PSG 71.7 %). In both scorers, Bland-Altman plots showed that the mean difference between sleep variables obtained from the two systems was within the acceptable range although the interclass correlations between the two systems were lower for StageN2 or StageR in either scorer. Although the results suggest that the sufficient inter-scorer reliability can be obtained with the EEG/ EOG recording system based on manual sleep stage scoring in healthy young adults, the nature of the EEG/ EOG recording system can influence the precisions in manually assessing sleep architectures in comparison to PSG.
机译:在临床和非临床环境中评估睡眠的自适应方法越来越多。本研究旨在调查睡眠阶段评分的分频性可靠性以及便携式脑电图(EEG)和电神学(EOG)记录系统中的睡眠变量评估。睡眠录音同时进行多重创新(PSG)和14名健康成年人的便携式EEG / EOG录音。在两种记录方法中评估了睡眠阶段评分的帧间互相协调。在每个分光器中使用Bland-Altman Plots的两种方法比较睡眠变量。 EEG / EOG数据的睡眠阶段评分的整体帧间频道协调率为85.5%,COHEN的Kappa为0.80(P <0.001),相当于PSG数据(89.2%,0.85,P <0.001 )对于STAGEN1的情况相对较低(EEG / EOG 60.1%,PSG 71.7%)。在这两个读得仪中,Bland-Altman图表明,从两个系统获得的睡眠变量之间的平均差异在可接受的范围内,尽管两个系统之间的次数相关性较低,用于在任一读射器中的STAGEN2或STAGER较低。虽然结果表明,通过基于健康年轻成年人的手动睡眠阶段评分的EEG / EOG记录系统可以获得足够的分光器可靠性,但EEG / EOG记录系统的性质可以在手动评估睡眠架构中影响精度与PSG相比。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号