There Is No Significant Difference in Fretting and Corrosion at the Trunnion of Metal and Ceramic Heads

机译:金属和陶瓷头耳期的烦恼和腐蚀没有显着差异

获取原文
获取原文并翻译 | 示例
           

摘要

No study has compared the prevalence of fretting and corrosion at the trunnion of cobalt-chrome (Co-Cr) and zirconia ceramic heads in the same patients. The authors sought to compare the prevalence of fretting and corrosion at the trunnion after total hip arthroplasty with a 32-mm Co-Cr or a 32-mm zirconia ceramic head-on-polyethylene bearing. Isolated revision of the acetabular components was performed in 78 patients (156 hips) for polyethylene wear and osteolysis of the acetabulum. Seventy-eight Co-Cr head-titanium alloy stem pairs were compared with 78 zirconia ceramic head-titanium alloy stem pairs in the same patients. Using a visual scoring technique, the 156 head-stem pairs were analyzed for the prevalence of fretting and/or corrosion at the trunnion of the stem and the prevalence of metal transfer to the inner taper of the femoral head. Mean patient age was 48 years, and there were 65 men and 13 women. There was no trunnionosis in either group. Prevalence of fretting (81% vs 83%) and corrosion (4% vs 3%) at the trunnion was not significantly different (P=. 518 vs.612, respectively) between the 2 groups. The median fretting scores (1.4 +/- 1.1 vs 1.2 +/- 1.4 points) and corrosion scores (1.2 +/- 0.8 vs 1.3 +/- 0.9 points) at the trunnion were not significantly different (P=. 893 vs.781, respectively) between the 2 groups. Prevalence of metal transfer to the inner taper of the femoral head (8% vs 10%) and the median metal transfer scores (1.2 +/- 0.9 vs 1.4 +/- 1.1 points) were not significantly different (P=. 213 vs.729, respectively) between the 2 groups.
机译:在同一患者中比较了在钴 - 铬(CO-CR)和氧化锆陶瓷头的耳轴上的微动和腐蚀的患病率。作者试图在总髋关节关节置换术和32毫米CO-Cr或32mM氧化锆陶瓷头 - 聚乙烯轴承中比较耳膜术中的搏动和腐蚀的患病率。髋臼组分的分离修改是在78名患者(156髋)中进行的聚乙烯磨损和髋臼骨溶解。将七十八种CO-CR头钛合金茎对与同一患者的78氧化锆陶瓷头钛合金茎对进行比较。使用视觉评分技术,分析156个头部茎对的茎干和/或腐蚀的患病率,以及金属转移到股骨头的内锥度的普遍性。平均患者年龄为48岁,男性和13名女性有65名。两组中没有耳鼻菌病。在2组之间的耳轴上的微动(81%vs 83%)和腐蚀(4%vs 3%)的腐蚀(4%vs 3%)在2组之间没有显着差异(P = 518 Vs.612)。在耳期的中位数分数(1.4 +/- 1.1 vs 1.2 +/- 1.4点)和腐蚀分数(1.2 +/- 0.8 vs 1.3 +/- 0.9点)没有显着差异(p = 893 vs.781分别在2组之间。金属转移到股骨头内锥度(8%vs 10%)和中值金属转移分数(1.2 +/- 0.9 Vs 1.4 +/- 1.1点)没有显着差异(P =。213 Vs. 729分别在2组之间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号