Reference materials of SO2, NO2 and CO were used to study the effects of different concentrations of NO2 and CO on the determination value of SO2 by flue gas analyzer with electrochemical sensors and UV differential optical absorption spectroscopy (DOAS). Results showed that SO2 was affected more greatly by NO2 than CO using the method of electrochemical sensors. The electrolytic reaction of NO2 was the primary factor causing the lowerconcentration of SO2. SO2 with the concentration of 0-300 mg/m(3) was affected greatly by NO2, and the determination value was unreliable. The determination error of SO2, with the concentration higher than 1500 mg/m(3), was about 2%, and the test value was more accurate. The method of UV DOAS could improve the determination reliability of concentration of SO2 greatly when NO2 coexisted.
机译:SO2,NO2和CO的参考材料用于研究具有电化学传感器和UV差分光学吸收光谱(DOA)的烟道气分析仪的不同浓度NO2和CO对SO2测定值的影响。 结果表明,使用电化学传感器的方法,SO2比CO2更大的影响。 NO2的电解反应是导致SO2的较低浓缩的主要因素。 浓度为0-300mg / m(3)的SO 2受到NO2的影响,测定值不可靠。 浓度高于1500mg / m(3)的SO2的测定误差约为2%,并且测试值更准确。 当No2共存时,UV DOA的方法可以大大提高SO2浓度的确定可靠性。