首页> 外文期刊>Magnetic resonance in medicine: official journal of the Society of Magnetic Resonance in Medicine >Method of B B 0 mapping with magnitude‐based correction for bipolar two‐point D D ixon cardiac MRI MRI
【24h】

Method of B B 0 mapping with magnitude‐based correction for bipolar two‐point D D ixon cardiac MRI MRI

机译:B B 0映射与基于幅度的双极双点D D D IXON心脏MRI MRI的方法

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose The conventional two‐point (2pt) Dixon technique explicitly estimates B0 map by performing phase unwrapping. When signal loss, phase singularity, artifacts, or spatially isolated regions corrupt the measured phase images, this unwrapping‐based technique will face difficulty. This work aims to improve the reliability of B0 mapping by performing unwrapping error correction. Method To detect the unwrapping‐caused phase errors, we determined a magnitude‐based fat/water mask and used it as reference to identify pixels being mismatched by the phase‐based mask, which was derived from the B0‐corrected phase term of the Hermitian product between echoes. Then, we corrected the afore‐determined phase error on a region‐by‐region basis. We tested the developed method with nine patients’ data, and the results were compared with a well‐established region‐growing technique. Results By adding the step to correct unwrapping‐caused error, we improved the robustness of B0 mapping, resulting in better fat–water separation when compared with the conventional 2pt and the phasor‐based region‐growing techniques. Conclusion We showed the feasibility of B0 mapping with bipolar 2pt human cardiac data. The software is freely available to the scientific community. Magn Reson Med 78:1862–1869, 2017. ? 2016 International Society for Magnetic Resonance in Medicine.
机译:目的通过执行阶段展开,传统的两点(2pt)迪克松技术明确估计B0映射。当信号丢失,相位奇点,伪像或空间隔离区域破坏测量的相位图像时,这种基于展开的技术将面临难度。这项工作旨在通过执行展开纠错来提高B0映射的可靠性。检测展开引起的相位误差的方法,我们确定了一种基于幅度的脂肪/水掩模,并使用它作为参考,以识别由基于相位的掩模不匹配的像素,这是源自隐士的B0校正相位术语回声之间的产品。然后,我们在逐个区域的基础上校正了上述相位误差。我们测试了具有九个患者数据的开发方法,并将结果与​​成熟的区域越来越多的技术进行了比较。结果通过添加步骤来纠正展开造成的误差,提高了B0映射的鲁棒性,导致与传统的2PT和基于相量的区域生长技术相比的更好的脂肪分离。结论我们展示了B0映射与双极2PT人体心脏数据的可行性。该软件自由地提供给科学界。 Magn Reson Med 78:1862-1869,2017 2016年医学中的磁共振学会。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号