Comparison of a New Swept-Source Anterior Segment Optical Coherence Tomography and a Scheimpflug Camera for Measurement of Corneal Curvature

机译:新扫源前段光学相干断层扫描和用于测量角膜曲率的Scheimpflug相机的比较

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: To evaluate the agreement of corneal curvature measured by a newly introduced swept-source anterior segment optical coherence tomography (CASIA2) and Pentacam (Oculus, Wetzlar, Germany). Methods: Corneal curvature was measured in 49 healthy subjects (age: 24.8 +/- 4.36) by using CASIA2 and Pentacam. Repeatability was evaluated for CASIA2, and agreement between the CASIA2 and Pentacam was assessed. The coefficient of repeatability, the relative coefficient of repeatability, the range of the limits of agreement were analyzed, and Bland-Altman plots were performed to assess the interdevice agreement of measurement. Results: For the intradevice reliability of the CASIA2, the mean difference between the 2 examinations was from 0.02 to 0.14 D for the corneal curvatures and the limits of agreement range was from 0.45 to 2.42 D for the corneal curvatures; there was no statistically significant difference found between the first and second examinations obtained from CASIA2 (P> 0.05). For agreement between the 2 devices, the Bland-Altman analysis of the posterior corneal curvature showed a mean difference of 0.16 D (P= 0.047) in the flat posterior corneal curvature and 0.16 D (P= 0.046) in the steep posterior corneal curvature. For the measurement of anterior corneal curvature and true net power (TNP) with the Pentacam and CASIA2, there was no significant difference in the flat anterior corneal curvature (P= 0.585), steep anterior corneal curvature (P= 0.274), flat TNP (P= 0.685), and steep TNP (P= 0.447). Conclusions: Our study indicated generally good agreement of corneal curvature measurement by using Pentacam and CASIA2. We consequently expect no difference in clinical outcome when measurements for corneal curvature, especially anterior corneal curvature and TNP, are performed with either of the devices.
机译:目的:评估通过新引入的扫掠源前段光学相干性断层扫描(CASIA2)和Pentacam(Oculus,Wetzlar,德国)测量的角膜曲率协议。方法:使用CASIA2和戊克酸克酸克,在49个健康受试者(年龄:24.8 +/- 4.36)中测量角膜曲率。对CASIA2评估重复性,评估CASIA2和戊酰胺之间的协议。分析重复性系数,相应的重复系数,分析了一致性限制的范围,并进行了平坦的地图,以评估测量的互脱协议。结果:对于CACEA2的内部内心可靠性,2个检查的平均差异为角膜曲率为0.02至0.14d,并且协议范围的限制为角膜曲率为0.45至2.42 d;从CACAIA2获得的第一和第二次检查之间没有发现统计学上有显着差异(P> 0.05)。对于2个装置之间的一致性,后角膜曲率的平坦-Atmman分析显示出平坦的后角膜曲率下的0.16d(p = 0.047)的平均差异,并且在陡峭的后角膜曲率中为0.16d(p = 0.046)。对于用戊肟和CACAIA2测量前角膜曲率和真正的净功率(TNP),扁平前角膜曲率没有显着差异(P = 0.585),陡峭的前角膜曲率(P = 0.274),平坦的TNP( P = 0.685),陡峭TNP(P = 0.447)。结论:我们的研究表明,通过使用近萨克姆和CACAIA2,通常对角膜曲率测量的吻合一致。因此,当用任一器件进行角膜曲率测量,特别是前角膜曲率和TNP的测量时,我们期望临床结果没有差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号