首页> 外文期刊>麻酔 >A simple lightweight device for measurement of low hemoglobin concentrations: the Hemo Cue Low Hemoglobin Photometer
【24h】

A simple lightweight device for measurement of low hemoglobin concentrations: the Hemo Cue Low Hemoglobin Photometer

机译:一种简单的轻质装置,用于测量低血红蛋白浓度:Hemo提示低血红蛋白光度计

获取原文
获取原文并翻译 | 示例
           

摘要

To determine low hemoglobin concentrations in the washed fluid is useful for calculation of blood loss and that in the supernatant is also useful for evaluation of hemolysis during extracorporeal circulation or in the blood preparations. The HemoCue Low Hemoglobin Photometer is a new lightweight device, which is easily movable anywhere and is able to determine hemoglobin concentration in low ranges (0.03-3.00 g.dl-1). We investigated the accuracy and reliability of the HemoCue Low Hemoglobin Photometer. Blood samples were diluted with saline containing heparin into six tubes. The hemoglobin concentrations of each diluted sample were measured by the HemoCue Low Hemoglobin Photometer, and were compared with the calculated values. A high positive correlation was found between the values measured by the HemoCue Low Hemoglobin Photometer (Y) and the calculated values (X); Y = 1.047 X, r = 0.999, P < 0.0001. It took only 11-54 seconds (mean value: 31.4 seconds) for the measurement. It is concluded that the HemoCue Low Hemoglobin Photometer is a reliable and useful device for measurement of low hemoglobin concentration.
机译:为了确定洗涤的流体中的低血红蛋白浓度可用于计算失血,并且在上清液中也可用于评估体外循环期间的溶血或血液制剂。 Hemocue低血红蛋白光度计是一种新的轻质装置,可容易地移动任何地方,并且能够确定低范围内的血红蛋白浓度(0.03-3.00 g.dl-1)。我们调查了血管低血红蛋白光度计的准确性和可靠性。将血液样品用含肝素的盐稀释成六管。通过血管低血红蛋白光度计测量每个稀释样品的血红蛋白浓度,并与计算值进行比较。在血管低血红蛋白光度计(Y)测量的值与计算值(x)之间的值之间发现了高阳性相关性; Y = 1.047 x,r = 0.999,p <0.0001。测量只花了11-54秒(平均值:31.4秒)。结论是,血压低血红蛋白光度计是用于测量低血红蛋白浓度的可靠且有用的装置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号