Calibration and validation of a MCC/IMS prototype for exhaled propofol online measurement

机译:校准和验证MCC / IMS原型进行呼出的异丙酚在线测量

获取原文
获取原文并翻译 | 示例
           

摘要

Propofol is a commonly used intravenous general anesthetic. Multi-capillary column (MCC) coupled Ion mobility spectrometry (IMS) can be used to quantify exhaled propofol, and thus estimate plasma drug concentration. Here, we present results of the calibration and analytical validation of a MCC/IMS pre market prototype for propofol quantification in exhaled air. Calibration with a reference gas generator yielded an R-2 >= 0.99 with a linear array for the calibration curve from 0 to 20 ppb(v). The limit of quantification was 0.3 ppb(v), and the limit of detection was 0.1 ppb(v). The device is able to distinguish concentration differences >0.5 ppb(v), for the concentration range between 2 and 4 ppb(v), and >0.9 ppb(v) for the range between 28 and 30 ppb(v). The imprecision at 20 ppb(v) is 11.3% whereas it is 3.5% at a concentration of 40 ppb(v). The carry-over duration is 3 min. The MCC/IMS we tested provided online quantification of gaseous propofol over the clinically relevant range at measurement frequencies of one measurement each minute. (C) 2017 Elsevier B.V. All rights reserved.
机译:异丙酚是一种常用的静脉内一般麻醉剂。多毛细管柱(MCC)耦合离子迁移率光谱(IMS)可用于量化呼出的异丙酚,从而估计血浆药物浓度。在这里,我们呈现MCC / IMS Pre Mass原型的校准和分析验证的结果,以在呼出的空气中进行异丙酚量化。使用参考气体发生器的校准产生R-2> = 0.99,具有从0到20 ppb(V)的校准曲线的线性阵列。定量限为0.3ppb(v),检测限为0.1ppb(v)。该装置能够区分浓度差异> 0.5ppb(v),在2至4ppb(v)之间的浓度范围,> 0.9ppb(v)的范围为28至30ppb(v)。 20 ppb(v)的不精确是11.3%,而其浓度为40 ppb(v)。随身持续时间为3分钟。我们测试的MCC / IMS在每分钟测量频率的测量频率下在临床相关范围内提供了在线定量的气态异丙酚。 (c)2017年Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号