Purpose: To introduce a novel multi‐source rotating shield brachytherapy (RSBT) apparatus for the precise simultaneous angular and linear positioning of all partially‐shielded 153Gd radiation sources in interstitial needles for treating prostate cancer. The mechanism is designed to lower the detrimental dose to healthy tissues, the urethra in particular, relative to conventional high‐dose‐rate brachytherapy (HDR‐BT) techniques. Methods: Following needle implantation, the delivery system is docked to the patient template. Each needle is coupled to a multi‐source afterloader catheter by a connector passing through a shaft. The shafts are rotated by translating a moving template between two stationary templates. Shaft walls as well as moving template holes are threaded such that the resistive friction produced between the two parts exerts enough force on the shafts to bring about the rotation. Rotation of the shaft is then transmitted to the shielded source via several keys. Thus, shaft angular position is fully correlated with the position of the moving template. The catheter angles are simultaneously incremented throughout treatment as needed, and only a single 360° rotation of all catheters is needed for a full treatment. For each rotation angle, source depth in each needle is controlled by a multi‐source afterloader, which is proposed as an array of belt‐driven linear actuators, each of which drives a source wire. Results: Optimized treatment plans based on Monte Carlo dose calculations demonstrated RSBT with the proposed apparatus reduced urethral D 1cc below that of conventional HDR‐BT by 35% for urethral dose gradient volume within 3 mm of the urethra surface. Treatment time to deliver 20 Gy with multi‐source RSBT apparatus using nineteen 62.4 GBq 153 Gd sources is 117 min. Conclusions: The proposed RSBT delivery apparatus in conjunction with multiple nitinol catheter‐mounted platinum‐shielded 153 Gd sources enables a mechanically feasible urethra‐sparing treatment technique for prostate cancer in a clinically reasonable timeframe.
机译:目的:介绍一种新颖的多源旋转屏蔽近距离放射治疗(RSBT)装置对于所有的精确同步角度和线性定位部分屏蔽153Gd辐射源在用于治疗前列腺癌间隙针。该机构被设计成降低有害的剂量对健康组织,特别是相对于传统的高剂量率近距离(HDR-BT)技术尿道。方法:遵循针注入,输送系统对接到病人模板。每个针通过穿过轴连接器耦合到多源后加载器导管。该轴通过翻译两个固定模板之间移动的模板旋转。轴壁以及移动模板孔是带螺纹的,使得这两个部件施加足够的力之间产生在轴带来的旋转电阻摩擦。所述轴的旋转,然后通过几个密钥发送到屏蔽源。因此,轴的角位置完全与移动模板的位置相关。根据需要将导管角度同时在整个治疗递增,并且需要用于一个完整的治疗仅所有导管的单个360°旋转。对于每个旋转角度,在各针源深度由多源后加载器,其被提议作为皮带驱动线性致动器的阵列,其中每一个驱动源线控制。结果:基于蒙特卡罗优化的治疗计划的剂量表现出RSBT与所提出的装置的计算尿道表面不到3mm降低至低于常规HDR-BT的尿道d 1cc的35%,持续尿道剂量梯度体积。处理时间以提供20戈瑞使用19 62.4吉贝153的Gd源多源RSBT装置是117分钟。结论:在与多个镍钛诺导管安装铂屏蔽153个的Gd源结合所提出的RSBT输送装置能够用于前列腺癌的机械可行尿道节约处理技术在临床上合理的时间框架。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号