In this study, the effect of column overload of the matrix ion, chloride, on the elution peak profiles of trace bromate is investigated. The resultant peak profiles of chloride and bromate are explained on the basis of competitive Langmuir isotherms. The Thermo IonPac AS9-HC, AS19 and AS23 columns are recommended by the manufacturer for bromate (a carcinogen) analysis. Under trace conditions, these columns provide baseline resolution of bromate from matrix ions such as chloride (R-s = 2.9, 3.3 and 3.2, respectively for the three columns). Injection of 10-300mM chloride with both hydroxide and carbonate eluents resulted in overload on these columns. On the basis of competitive Langmuir isotherms, a deficiency in the local concentration of the more retained eluent in addition to analyte overload leads to fronting of the overloaded analyte peak. The peak asymmetries (B/A(10%)) for chloride changed from 1.0 (Gaussian) under trace conditions to 0.7 (fronting) at 300mM Cl- for lonPac AS9-HC, 0.9-0.6 for AS19 and 0.8-0.5, for AS23, respectively. The 10mM bromate peak is initially near Gaussian (B/A(10%) = 0.9) but becomes increasingly distorted and pulled back into the chloride peak as the concentration of chloride increased. Increasing the eluent strength reduced the pull-back effect on bromate and fronting in chloride in all cases. (C) 2016 Elsevier B.V. All rights reserved.
机译:在这项研究中,研究了基质离子氯离子的色谱柱超载对痕量溴酸盐洗脱峰分布的影响。根据竞争性的朗缪尔等温线解释了所得的氯化物和溴酸盐的峰形。制造商建议使用Thermo IonPac AS9-HC,AS19和AS23色谱柱进行溴酸盐(致癌物)分析。在痕量条件下,这些色谱柱提供了基线离子(例如三根色谱柱的Rs = 2.9、3.3和3.2)从基质离子(如氯离子)中分离出溴酸盐的基线分辨率。注入10-300mM氯化物以及氢氧化物和碳酸盐洗脱液会导致这些色谱柱超载。根据竞争性Langmuir等温线,除了分析物过载之外,保留更多的洗脱液的局部浓度不足会导致分析物峰值超前。氯的峰不对称度(B / A(10%))从迹线条件下的1.0(高斯)变为lonPac AS9-HC在300mM Cl-下为0.7(前沿),AS19为0.9-0.6,AS23为0.8-0.5 , 分别。溴酸盐的10mM峰最初接近高斯(B / A(10%)= 0.9),但随着氯离子浓度的增加,其畸变越来越多,并拉回到氯离子峰中。在所有情况下,提高洗脱液强度均会降低对溴酸盐的拉回作用和在氯中的前沿。 (C)2016 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号