首页> 外文期刊>Journal of Agricultural and Food Chemistry >Performance assessment and validation of a paramagnetic particle-based enzyme-linked immunosorbent assay for chlorpyrifos in agricultural runoff waters
【24h】

Performance assessment and validation of a paramagnetic particle-based enzyme-linked immunosorbent assay for chlorpyrifos in agricultural runoff waters

机译:基于顺磁性颗粒的酶联免疫吸附法测定农业径流水中毒死rif的性能评估和验证

获取原文
获取原文并翻译 | 示例
           

摘要

A commercial magnetic particle-based enzyme-linked immunosorbent assay (ELISA) kit for the insecticide chlorpyrifos [O,O-diethyl-O-(3,5,6-trichloro-2-pyridyl) phosphorothioate] was evaluated for its specificity, precision, and accuracy, its susceptibility to matrix interferences in agricultural and environmental surface waters, and its comparability to a gas chromatographic/flame photometric (GC/FPD) method for the determination of organophosphorus pesticides in natural waters. Repeatability, reproducibility, and accuracy studies show that the kit satisfies current U.S. Environmental Protection Agency criteria for the assessment of analytical methods. Observable matrix effects were found to be present in all of the environmental test waters, with the slopes of calibration curves generated in each of the test matrices deviating from that of the control matrix by as much as 16%. Specificity studies indicate that the chlorpyrifos polyclonal antibody adequately differentiates the target compound from other structurally similar organophosphorus pesticides, with the exception of its methyl analogue. Cross-reactivity with chlorpyrifos-methyl was approximately 37%, while reactivity with diazinon, pyridaphenthion, diclofenthion, bromiphos-ethyl, bromiphos-methyl, pirimiphos-ethyl, and chlorpyrifos oxon ranged from 1.6 to 10.7%. Cross-reactivity with pirimiphos-methyl, 3,5,6-trichloro-2-pyridinol, diethyl phosphate, and diethyl thiophosphate was negligible (<1%). Validation of the paramagnetic particle ELISA format was accomplished using water samples from two monitoring studies that were collected, split, and analyzed directly by ELISA and by GC/FPD. Results of the two analytical methods were then compared using standard t tests, regression analysis, and differences against mean measurement (bias) plots. While the agreement between the two methods was determined to be satisfactory, ELISA exhibits consistent positive bias in environmental matrices. Several preanalysis mitigation steps were suggested that may help moderate bias, but additional study is recommended to explicate the exact factors responsible for its consistent overestimation of results. Keywords: ELISA; chlorpyrifos; paramagnetic particles; performance evaluation; matrix effects; comparative validation.
机译:评价了一种用于杀虫剂毒死rif [O,O-二乙基-O-(3,5,6-三氯-2-吡啶基)硫代磷酸酯]的基于商业磁性颗粒的酶联免疫吸附测定(ELISA)试剂盒的特异性,精密度,准确度,对农业和环境地表水中基质干扰的敏感性以及与气相色谱/火焰光度法(GC / FPD)测定自然水中有机磷农药的方法的可比性。重复性,可重复性和准确性研究表明,该试剂盒符合美国环境保护署当前对分析方法进行评估的标准。发现在所有环境测试水中均存在可观察到的基质效应,每个测试矩阵中生成的校准曲线的斜率与对照矩阵的斜率相差多达16%。特异性研究表明,毒死rif多克隆抗体可将目标化合物与其他结构相似的有机磷农药(甲基类似物除外)充分区分开。与毒死rif甲基的交叉反应性约为37%,而与二嗪农,哒嗪酮,双氯芬硫磷,溴代磷乙基,溴代磷甲基,吡imi磷和乙基毒死的反应性为1.6%至10.7%。与嘧啶磷甲基,3,5,6-三氯-2-吡啶醇,磷酸二乙酯和硫代磷酸二乙酯的交叉反应可忽略不计(<1%)。顺磁粒子ELISA格式的验证是使用来自两个监测研究的水样品完成的,这些样品直接通过ELISA和GC / FPD收集,分离和分析。然后使用标准t检验,回归分析和相对于均值测量(偏差)图的差异比较了两种分析方法的结果。虽然确定两种方法之间的一致性令人满意,但ELISA在环境基质中显示出一致的正偏差。建议采取一些缓解分析前的步骤,可能有助于缓解偏见,但建议进行其他研究,以阐明造成其一贯高估结果的确切因素。关键字:ELISA;毒死rif;顺磁性粒子绩效评估;基质效应比较验证。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号