Comparison of central corneal thickness measurements by ultrasound pachymetry, konan noncontact optical pachymetry, and orbscan pachymetry.

机译:通过超声测厚仪,konan非接触式光学测厚仪和orbscan测厚仪比较中央角膜厚度。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To assess the repeatability of two optical techniques for measurement of central corneal thickness and to compare the validity of these techniques with standard ultrasound pachymetry. METHODS: Corneal thickness was measured twice with Konan optical pachymetry, Orbscan optical pachymetry, and 3 times by conventional ultrasound pachymetry in 100 eyes of 50 healthy volunteers between the ages of 22 and 74 years. The repeatability of measurements of each instrument was determined and compared. Validity was determined by comparing the means of the first Konan and Orbscan measurements to the mean of standard ultrasound pachymetry. Intraclass correlation coefficients (ICCs) were calculated, and their confidence intervals (CIs) were determined. RESULTS: Repeatability of ultrasound pachymetry had an ICC of 0.905 (95% CI: 0.870-0.932). The Konan pachymeter had an ICC of 0.947 (95% CI: 0.922-0.964), and the Orbscan pachymetry yielded an ICC of 0.972 (95% CI: 0.959-0.981). Validity of Konan pachymetry hadan ICC of 0.730 (95% CI: 0.623-0.810). Orbscan pachymetry had an ICC of 0.865 (95% CI: 0.805-0.907). CONCLUSIONS: Both of the newer optical methods for determining corneal thickness produced measurements of greater repeatability than the standard ultrasound technique. Orbscan pachymetry provides greater repeatability (P = 0.02) and validity (P = 0.007) than the Konan optical pachymeter.
机译:目的:评估两种用于测量中央角膜厚度的光学技术的可重复性,并将这些技术与标准超声测厚法进行比较。方法:用Konan光学测厚仪,Orbscan光学测厚仪两次测量角膜厚度,常规超声测厚仪测量50例22岁至74岁健康志愿者的100只眼中的3次。确定并比较每种仪器的测量重复性。通过将第一次Konan和Orbscan测量的平均值与标准超声测厚仪的平均值进行比较来确定有效性。计算类内相关系数(ICC),并确定其置信区间(CIs)。结果:超声测厚法的可重复性具有0.905的ICC(95%CI:0.870-0.932)。 Konan测厚仪的ICC为0.947(95%CI:0.922-0.964),而Orbscan测厚仪的ICC为0.972(95%CI:0.959-0.981)。 Konan测厚法的有效值具有0.730(95%CI:0.623-0.810)。 Orbscan测厚法的ICC为0.865(95%CI:0.805-0.907)。结论:两种较新的光学方法用于确定角膜厚度均比标准超声技术具有更高的可重复性。与Konan光学测厚仪相比,Orbscan测厚仪具有更高的可重复性(P = 0.02)和有效性(P = 0.007)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号