首页> 外文期刊>Journal of pediatric rehabilitation medicine >Measuring reliability and validity of the ActiGraph GT3X accelerometer for children with cerebral palsy: A feasibility study
【24h】

Measuring reliability and validity of the ActiGraph GT3X accelerometer for children with cerebral palsy: A feasibility study

机译:ActiGraph GT3X加速度计在脑瘫儿童中的信度和效度测量:一项可行性研究

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: The purposes of this study were to: 1) establish inter-instrument reliability between left and right hip accelerometer placement; 2) examine procedural reliability of a walking protocol used to measure physical activity (PA); and 3) confirm concurrent validity of accelerometers in measuring PA intensity as compared to the gold standard of oxygen consumption measured by indirect calorimetry. METHODS: Eight children (mean age: 11.9; SD: 3.2, 75% male) with CP (GMFCS levels I-III) wore ActiGraph GT3X accelerometers on each hip and the Cosmed K4b2portable indirect calorimeter during two measurement sessions in which they performed the six minute walk test (6MWT) at three self-selected speeds (comfortable/slow, brisk, fast). Oxygen consumption (VO2) and accelerometer step and activity count data were recorded. RESULTS: Inter-instrument reliability of ActiGraph GT3X accelerometers placed on left and right hips was excellent (ICC=0.96-0.99, CI95: 0.81-0.99). Reproducibility of the protocol was good/excellent (ICC=0.75-0.95, CI95: 0.75-0.98). Concurrent validity of accelerometer count data and VO2 was fair/good (rho=0.67, p< 0.001). The correlation between step count and VO2 was not significant (rho=0.29, p=0.2). CONCLUSION: This preliminary research suggests that ActiGraph GT3X accelerometers are reliable and valid devices to monitor PA during walking in children with CP and may be appropriate in rehabilitation research and clinical practice. ActiGraph GTX3 step counts were not valid for this sample and further research is warranted.
机译:目的:本研究的目的是:1)建立左右髋部加速度计放置之间的仪器间可靠性; 2)检查用于测量身体活动(PA)的步行方案的程序可靠性;和3)确认加速度计与通过间接量热法测得的耗氧金标准相比,在同时测量PA强度方面的有效性。方法:八名儿童(平均年龄:11.9; SD:3.2,75%男性)的CP(I-III级GMFCS)在每个髋关节上佩戴了ActiGraph GT3X加速度计,并在两次测量中分别使用了Cosmed K4b2便携式间接量热计,在其中进行了六个在三种自选速度下(舒适/慢速,轻快,快速)的分钟步行测试(6MWT)。记录耗氧量(VO2)和加速度计步数和活动计数数据。结果:放在左右臀部的ActiGraph GT3X加速度计的仪器间可靠性极好(ICC = 0.96-0.99,CI 95 :0.81-0.99)。该方案的重复性良好/优异(ICC = 0.75-0.95,CI 95 :0.75-0.98)。加速度计计数数据和VO2的并发有效性是公平/良好(rho = 0.67,p <0.001)。步数与VO2之间的相关性不显着(rho = 0.29,p = 0.2)。结论:这项初步研究表明,ActiGraph GT3X加速度计是可靠且有效的设备,可在CP儿童步行期间监测PA,并且可能适合进行康复研究和临床实践。 ActiGraph GTX3步数对此样品无效,因此有必要进行进一步研究。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号