Elimination of autosampler carryover in a bioanalytical HPLC-MS/MS method: a case study.

机译:在生物分析HPLC-MS / MS方法中消除自动进样器残留:一个案例研究。

获取原文
获取原文并翻译 | 示例
           

摘要

A case study in identifying and eliminating the source of autosampler carryover in a bioanalytical HPLC-MS/MS assay is described. Through a series of systematic experiments, the carryover was traced to the injection valve and was eliminated by switching from a partial loop to a full loop injection, which provided more effective flushing of the sample flow path. The susceptibility of the HPLC system to carryover was demonstrated to depend on the absolute sensitivity of the detection method and the mass of analyte injected at the assay lower limit of quantitation (LLOQ).
机译:描述了在生物分析HPLC-MS / MS分析中识别和消除自动进样器残留来源的案例研究。通过一系列系统实验,将残留物追溯到进样阀,并通过从部分定量进样切换到全定量进样来消除残留物,从而更有效地冲洗了样品流路。 HPLC系统对残留物的敏感性被证明取决于检测方法的绝对灵敏度和在测定下限(LLOQ)下注入的分析物的质量。