Discarded ProFile and ProTaper nickel-titanium rotary instruments, with unknown history of clinical use, were obtained from graduate endodontic clinics at Ohio State University and University of Texas Health Science Center at Houston Dental Branch. These discarded instruments and as-received instruments of both types were examined with a scanning electron microscope to investigate effects of clinical use and causes of failure. For used ProTaper instruments, dentinal debris was wedged mostly in narrow, radial, land-type regions and less on convex flute surfaces. For used ProFile instruments, dentinal debris was wedged mostly in the metal rollover and on concave flute surfaces. Used instruments of both types exhibited widened machining grooves, and elongated and stretched roll-over. Dentin chips were wedged in surface micro-cracks that appeared to propagate from original machining flaws and widen during in vivo root canal preparation. From our observational study, wedged dentinal deposits seem to play a pivotal role for clinical failure of these instruments.
机译:废弃的ProFile和ProTaper镍钛旋转器械是从俄亥俄州立大学的研究生牙髓诊所和德克萨斯大学休斯敦牙科分校的健康科学中心的牙髓诊所获得的,其临床使用历史未知。用扫描电子显微镜检查了这两种丢弃的仪器和已接收的仪器,以研究临床使用的效果和失败的原因。对于使用过的ProTaper器械,牙本质碎屑大多楔入狭窄的径向陆地型区域,而楔形碎屑表面则较少。对于使用过的ProFile仪器,牙本质碎屑大多楔在金属翻转表面和凹形凹槽表面上。两种类型的二手仪器都显示出加宽的加工槽,以及伸长和拉伸的翻转。牙本质芯片被楔入表面微裂纹中,表面微裂纹似乎是从原始的加工缺陷中传播出来的,并在体内根管准备过程中变宽。根据我们的观察性研究,楔状牙本质沉积物似乎在这些器械的临床失败中起着关键作用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号