Influence of instrument size on the accuracy of different apex locators: an in vitro study.

机译:器械尺寸对不同根尖定位器精度的影响:一项体外研究。

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of this in vitro investigation was to determine the accuracy of 4 different electronic apex locators (EALs) with 3 different instrument sizes. For this study 146 roots were embedded in an agar solution. Electronic measurements were made to the physiologic foramen (apical constriction) with the Elements Apex Locator, Justy II, Raypex 5, and ProPex II and K-type files sizes 08, 10, and 15. Statistical significances were calculated with the sign test (P < .001). Exact measurements to the physiologic foramen were made with the Elements Apex Locator, 36.99%, 39.04%, and 44.93%; Justy II, 38.62%, 32.41%, and 43.41%; Raypex 5, 42.76%, 39.31%, and 39.06%; and ProPex II, 38.62%, 43.45%, and 40.63% of the time with instrument sizes 08, 10, and 15, respectively. No significant differences were found between the actual working length and EALs/instrument size. A nonsignificant higher number of unstable measurements were observed in all EALs with instrument size 15.
机译:这项体外研究的目的是确定具有3种不同仪器尺寸的4种不同电子根尖定位器(EAL)的准确性。对于本研究,将146个根嵌入琼脂溶液中。使用Elements Apex Locator,Justy II,Raypex 5和ProPex II以及K型文件分别为08、10和15的尺寸对生理孔(根部狭窄)进行电子测量。 <.001)。使用Elements Apex Locator对生理学孔进行了精确测量,分别为36.99%,39.04%和44.93%; Justy II,38.62%,32.41%和43.41%; Raypex 5、42.76%,39.31%和39.06%;和ProPex II,仪器尺寸分别为08、10和15的时间分别为38.62%,43.45%和40.63%。在实际工作时间和EAL /仪器尺寸之间没有发现显着差异。在仪器尺寸为15的所有EAL中,观察到的不稳定测量值均没有明显增加。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号