High sensitivity analysis of oxprenolol in urine by capillary electrophoresis with C 18 packed on-line preconcentrator

机译:装有C 18在线预浓缩器的毛细管电泳高灵敏度分析尿中的羟戊诺泊酚

获取原文
获取原文并翻译 | 示例
           

摘要

High sensitivity analysis of oxprenolol in spiked human urine has been performed by capillary zone electrophoresis (CZE) in ammonium formate buffer pH 2.5 using an uncoated capillary with 1 cm length C 18 on-capillary preconcentrator at the inlet side. The preconcentrator was fabricated in laboratory using the packing method and not encapped C18 5 mu m particles as stationary phase material. The packed path was retained into the capillary by sintered stationary phase frits. Before running the CZE analysis, the oxprenolol was eluted from the preconcentrator by injecting a short plug of acetonitrile/water mixtures. With respect to classical CZE, the use of on-line preconcentrator widely increased the method sensitivity allowing the detection of the drug at 0.5 ng/mL (injected concentration). The method showed a linear response in the range of 1-150 ng/mL oxprenolol standard compound. The intra-day repeatability (n = 11) R.S.D. values for migration time, peak area and normalized peak area were 0.72%, 3.96% and 3.66%, respectively, while inter-day repeatability (n = 5 days) R.S.D. values were 2.74%, 9.41% and 9.83%, respectively. The method was successfully applied to the analysis of oxprenolol in extracted urine spiked at 250 pg/mL (oxprenolol LOQ concentration in urine). (c) 2005 Elsevier B.V. All rights reserved.
机译:毛细管区带电泳(CZE)在甲酸铵缓冲液pH 2.5中进行了毛细管区带电泳(CZE),使用毛细管长度为C 18的未包被毛细管在进口预浓缩器上进行了高灵敏度分析。该预浓缩器是在实验室中使用填充方法制造的,未包封5微米的C18颗粒作为固定相材料。通过烧结固定相玻璃料将填充路径保留在毛细管中。在进行CZE分析之前,通过注入一小段乙腈/水混合物塞子,从预浓缩器中洗脱出氯丁诺尔。对于传统的CZE,在线预浓缩器的使用大大提高了方法的灵敏度,可检测0.5 ng / mL(注射浓度)的药物。该方法显示的线性响应范围为1-150 ng / mL的氧苯丙醇标准化合物。日内重复性(n = 11)R.S.D.迁移时间,峰面积和归一化峰面积的值分别为0.72%,3.96%和3.66%,而日间重复性(n = 5天)R.S.D.分别为2.74%,9.41%和9.83%。该方法已成功地应用于加标为250 pg / mL的提取尿液中尿液中的oxprenolol分析(尿中的oxprenol LOQ浓度)。 (c)2005 Elsevier B.V.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号