首页> 外文期刊>Journal of chromatography, B. Analytical technologies in the biomedical and life sciences >Determination of total iodine in serum and urine samples by ion chromatography with pulsed amperometric detection - Studies on analyte loss, optimization of sample preparation procedures, and validation of analytical method
【24h】

Determination of total iodine in serum and urine samples by ion chromatography with pulsed amperometric detection - Studies on analyte loss, optimization of sample preparation procedures, and validation of analytical method

机译:脉冲安培检测-离子色谱法测定血清和尿液中的总碘-分析物损失,样品制备程序的优化和分析方法的验证

获取原文
获取原文并翻译 | 示例
           

摘要

A fast, accurate and precise ion chromatography method with pulsed amperometric detection was applied to evaluate a variety of parameters affecting the determination of total iodine in serum and urine of 81 subjects, including 56 obese and 25 healthy Polish children. The sample pretreatment methods were carried out in a closed system and with the assistance of microwaves. Both alkaline and acidic digestion procedures were developed and optimized to find the simplest combination of reagents and the appropriate parameters for digestion that would allow for the fastest, least time consuming and most cost-effective way of analysis. A good correlation between the certified and the measured concentrations was achieved. The best recoveries (96.8% for urine and 98.8% for serum samples) were achieved using 1 ml of 25% tetramethylammonium hydroxide solution within 6 min for 0.1 ml of serum/urine samples. Using 0.5 ml of 65% nitric acid solution the best recovery (95.3%) was obtained when 7 min of effective digestion time was used. Freeze-thaw stability and long-term stability were checked. After 24 weeks 14.7% loss of iodine in urine, and 10.9% in serum samples occurred. For urine samples, better correlation (R~2 = 0.9891) of various sample preparation procedures (alkaline digestion and application of OnGuard RP cartidges) was obtained. Significantly lower iodide content was found in samples taken from obese children. Serum iodine content in obese children was markedly variable in comparison with the healthy group, whereas the difference was less evident when urine samples were analyzed. The mean content in serum was 59.12 ± 8.86 μg/L, and in urine 98.26 ± 25.93 for obese children when samples were prepared by the use of optimized alkaline digestion reinforced by microwaves. In healthy children the mean content in serum was 82.58 ± 6.01 μg/L, and in urine 145.76 ±31.44 μg/L.
机译:快速,准确,精确的离子色谱法结合脉冲安培检测技术被用于评估影响包括81名肥胖者和25名波兰健康儿童在内的81名受试者血清和尿液中总碘测定的各种参数。样品预处理方法是在密闭系统中并借助微波进行的。开发了碱性和酸性消化程序并对其进行了优化,以找到最简单的试剂组合和适当的消化参数,从而实现最快,最省时,最经济高效的分析方法。在认证浓度和测量浓度之间实现了良好的相关性。对于0.1 ml的血清/尿液样品,在6分钟内使用1 ml的25%的四甲基氢氧化铵溶液可实现最佳的回收率(尿液为96.8%,血清样品为98.8%)。当使用7分钟的有效消化时间时,使用0.5 ml的65%硝酸溶液可获得最佳的回收率(95.3%)。检查了冻融稳定性和长期稳定性。 24周后,尿液中碘流失了14.7%,血清样品中碘流失了10.9%。对于尿液样品,各种样品制备程序(碱性消化和OnGuard RP软骨的应用)获得了更好的相关性(R〜2 = 0.9891)。肥胖儿童的样本中碘化物含量明显降低。与健康组相比,肥胖儿童的血清碘含量存在明显差异,而对尿液样本进行分析时差异并不明显。当使用微波强化的优化碱性消化法制备样品时,肥胖儿童的平均血清含量为59.12±8.86μg/ L,尿液的平均含量为98.26±25.93。在健康儿童中,血清平均含量为82.58±6.01μg/ L,尿液中的平均含量为145.76±31.44μg/ L。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号