...
首页> 外文期刊>Clinical chemistry and laboratory medicine: CCLM >Comparison of several point-of-care testing (POCT) glucometers with an established laboratory procedure for the diagnosis of type 2 diabetes using the discordance rate. A new statistical approach.
【24h】

Comparison of several point-of-care testing (POCT) glucometers with an established laboratory procedure for the diagnosis of type 2 diabetes using the discordance rate. A new statistical approach.

机译:使用不合格率将几种即时检验(POCT)血糖仪与已建立的实验室程序用于诊断2型糖尿病的比较。一种新的统计方法。

获取原文
获取原文并翻译 | 示例
           

摘要

The applicability of point-of-care testing (POCT) glucometers for monitoring blood glucose concentrations has been demonstrated. However, their use in diagnosing type 2 diabetes is still debated. Therefore, a new statistical procedure for estimating discordance rates (DRs) was applied in comparing a well-established laboratory method (Ebio) with another laboratory method (Cobas Integra 700) and with several POCT glucometers (Accu-Chek, Accutrend, Elite, HemoCue, Omni) in detecting glucose intolerance states. All procedures led to parallel glucose concentration patterns in capillary blood, venous plasma, and venous blood during oral glucose tolerance tests. However, the mean concentrations differed more or less. The Ebio and Integra results agreed within a maximal deviation of 3%. In blood samples, the HemoCue and Accutrend results were closest to the laboratory procedures (Ebio and Integra) and the highest differences were obtained with the Elite. Comparing whole blood values with those obtained in theaqueous blood compartment (Omni), even greater differences were observed. When all procedures were referred to the same glucose standard, the Ebio, Integra, Accutrend, and Omni results remained almost unchanged, whereas the Elite "moved" toward the Ebio results, and the Accu-Chek results toward the Omni results. Thus, traceability to an aqueous standard was observed with the Ebio, Integra, Accutrend, and Elite in all three sample systems. The Accu-Chek was only traceable in the presence of albumin, and HemoCue was not traceable at all. The clinical relevance of the differences observed between Ebio and POCT glucometers was tested by comparing the relative number of discordant classifications. The highest DRs were observed in the fasting state. They were higher in capillary blood than in the other sample systems. The DRs were found higher with POCT glucometers than with the other established laboratory procedure (Integra). Thus, at least in the fasting state, all POCT glucometers were less reliable than the established laboratory procedures and above the chosen criteria of clinical acceptability (DR < or = 5%). After transforming all glucometer results with a regression function (bias correction), the DRs were less than 5% if compared with the Ebio procedure in all sample systems. In conclusion, the WHO recommendation not to use POCT glucometers for diagnosing type 2 diabetes must be supported. However, after proper recalibration, the tested systems were acceptable. Therefore, manufacturers should reconsider their calibration procedure. Those POCT procedures should be preferred that can be referred to aqueous glucose solutions.
机译:已经证明了即时检验(POCT)血糖仪在监测血糖浓度方面的适用性。但是,它们在诊断2型糖尿病中的用途仍存在争议。因此,采用了一种新的估计不一致率(DR)的统计程序,以将一种完善的实验室方法(Ebio)与另一种实验室方法(Cobas Integra 700)以及几种POCT血糖仪(Accu-Chek,Accutrend,Elite,HemoCue)进行比较,Omni)检测葡萄糖不耐受状态。在口服葡萄糖耐量试验期间,所有程序均导致毛细血管血,静脉血浆和静脉血中葡萄糖浓度平行变化。但是,平均浓度或多或少地有所不同。 Ebio和Integra结果一致,最大偏差为3%。在血液样本中,HemoCue和Accutrend结果最接近实验室程序(Ebio和Integra),并且使用Elite获得的差异最大。将全血值与在水性血液隔室(Omni)中获得的值进行比较,可以观察到更大的差异。当所有程序都使用相同的葡萄糖标准时,Ebio,Integra,Accutrend和Omni的结果几乎保持不变,而Elite则“移向” Ebio的结果,而Accu-Chek的结果则转向Omni的结果。因此,在所有三个样品系统中,使用Ebio,Integra,Accutrend和Elite都可以观察到对水标样的可追溯性。 Accu-Chek仅在存在白蛋白的情况下可追溯,而HemoCue则根本无法追溯。通过比较不一致分类的相对数量,测试了Ebio和POCT血糖仪之间观察到的差异的临床相关性。在禁食状态下观察到最高的DR。它们在毛细血管血液中比在其他样品系统中更高。发现POCT血糖仪的DR高于其他已建立的实验室程序(Integra)。因此,至少在禁食状态下,所有POCT血糖仪的可靠性都低于既定的实验室程序,并且高于临床可接受性的选定标准(DR <或= 5%)。用回归函数(偏倚校正)转换所有血糖仪结果后,与所有样品系统中的Ebio方法相比,DR均小于5%。总之,必须支持WHO建议不要使用POCT血糖仪诊断2型糖尿病。但是,经过适当的重新校准后,被测系统是可以接受的。因此,制造商应重新考虑其校准程序。那些可以称为葡萄糖水溶液的POCT程序应该是优选的。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号