A pragmatic protocol for reduction in the metal artifact and radiation dose in multislice computed tomography of the spine: cadaveric evaluation after cervical pedicle screw placement.

机译:减少脊柱的多层计算机断层扫描中的金属伪影和辐射剂量的实用协议:颈椎椎弓根螺钉置入后的尸体评估。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To assess the correlation between the metal artifact degree and the tube voltage (in peak kilovolts; kVp) and tube current (in milliamperes; mA) in multidetector row computed tomography (MDCT) and evaluate the proper protocols for reduction in the metal artifact and the radiation dose. METHODS: We performed MDCT examinations for 13 cadavers operated for transpedicular screw insertion from C3 to C7, bilaterally. We used 80, 100, and 120 kVp in all cadavers. For the mA, we arbitrarily applied 100, 150, and 200 in 6 cadavers (experiment 1); 140, 180, and 220 in 4 (experiment 2), and randomly used mAs between 60 and 220 in 3 (experiment 3). The lengths of the artifacts emanating from the anterior and posterior tips of the screw were used for the degree of artifact. RESULTS: In experiment 1, the mA did not significantly affect the metal artifacts (P > 0.05); but the kVp did (P < 0.05). In experiment 2, the mA and kVp did not significantly affect the metal artifact degree (P > 0.05). In experiment3, the mA did not affect the metal artifacts, but the kVp caused changes in metal artifact degrees (simple quantitative analysis). CONCLUSIONS: The proper parameters for the simultaneous reduction in the artifact and radiation dose are suggested to be tube voltages from 100 to 120 kVp and tube currents lower than the generally used values of 200 to 220 mA.
机译:目的:评估多探测器行计算机断层扫描(MDCT)中金属伪影度与管电压(峰值千伏; kVp)和管电流(毫安; mA)之间的相关性,并评估减少金属伪影的适当方案和辐射剂量。方法:我们对13具尸体进行了MDCT检查,以进行双侧椎弓根螺钉从C3到C7的插入。我们在所有尸体中都使用了80、100和120 kVp。对于mA,我们在6个尸体中任意使用100、150和200(实验1); 140、180和220 in 4(实验2),并随机使用60和220 in 3(实验3)之间的mAs。从螺钉的前尖端和后尖端发出的伪影的长度用于伪影的程度。结果:在实验1中,mA没有显着影响金属伪影(P> 0.05);但kVp有(P <0.05)。在实验2中,mA和kVp对金属伪影度没有显着影响(P> 0.05)。在实验3中,mA不会影响金属伪影,但kVp会导致金属伪影度发生变化(简单的定量分析)。结论:同时减少伪影和辐射剂量的适当参数建议为100至120 kVp的管电压和低于200至220 mA的常用值的管电流。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号