首页> 外文期刊>Journal of computer assisted tomography >Fractional Quantitative Computed Tomography for Bone Mineral Density Evaluation: Accuracy, Precision, and Comparison to Quantitative Computed Tomography.
【24h】

Fractional Quantitative Computed Tomography for Bone Mineral Density Evaluation: Accuracy, Precision, and Comparison to Quantitative Computed Tomography.

机译:用于骨矿物质密度评估的分数定量计算机断层扫描:准确性,精确度以及与定量计算机断层扫描的比较。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE:: To evaluate bone mineral density considering its distribution, fractional quantitative computed tomography (fQCT) was designed and verified. METHODS:: Quantitative computed tomography (QCT) was performed at 64 areas in 10 swine long bones. Fractional quantitative computed tomography was measured at the identical areas as the proportion of pixels showing a bone density higher than 130 mg/mL equivalent. All target areas were extracted and incinerated to measure apparent ash bone density. Based on standard references, the accuracy and precision of fQCT were evaluated and the results were compared with conventional QCT results. RESULTS:: The correlation coefficient between fQCT and apparent ash bone density was 0.843 (P < 0.0001). The fQCT showed good correlation with volume fraction (r = 0.88, P < 0.0001). The coefficient of variation of fQCT was 0.42%. The fQCT revealed higher accuracy and precision than the results of QCT. CONCLUSION:: Fractional quantitative computed tomography was designedand verified as a reliable method to measure bone mineral density.
机译:目的:为了评估骨矿物质密度的分布,设计并验证了分数定量计算机断层扫描(fQCT)。方法:在10头猪长骨的64个区域进行了定量计算机断层扫描(QCT)。在相同区域测量分数定量计算机断层扫描,其显示的骨密度高于130 mg / mL当量的像素比例。提取并焚烧所有目标区域,以测量表观骨灰密度。基于标准参考,评估了fQCT的准确性和精密度,并将结果与​​常规QCT结果进行了比较。结果:fQCT与表观骨灰密度之间的相关系数为0.843(P <0.0001)。 fQCT与体积分数显示出良好的相关性(r = 0.88,P <0.0001)。 fQCT的变异系数为0.42%。 fQCT显示出比QCT结果更高的准确性和精确度。结论:分数定量计算机断层扫描被设计和验证为一种可靠的方法来测量骨矿物质密度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号