Accuracy of the CNAP monitor, a noninvasive continuous blood pressure device, in providing beat-to-beat blood pressure readings in the prone position

机译:CNAP监护仪(一种非侵入性连续血压设备)在俯卧位置提供逐次血压读数的准确性

获取原文
获取原文并翻译 | 示例
           

摘要

Study Objective To assess the accuracy of a noninvasive continuous arterial pressure (CNAP) monitor in patients who are positioned prone in the operating room. Design Prospective study. Setting Operating room at a children's hospital. Patients 20 pediatric patients, aged 13.8 ± 2 years, and weight 63.7 ± 18.8 kg, scheduled for surgery in the prone position, and for which arterial catheter placement was planned. Interventions Measurements were recorded with an arterial line (AL) and a new noninvasive continuous blood pressure (BP) monitor. Measurements Systolic (SBP), diastolic (DBP), and mean arterial (MAP) pressure readings were captured from an arterial cannula and the CNAP device every minute during anesthesia. Main Results The study cohort consisted of analysis of 4104 pairs of SBP, DBP, and MAP values, which showed an absolute difference between the AL and CNAP device readings of 7.9 ± 6.3 mmHg for SBP, 5.3 ± 4.3 mmHg for DBP, and 4.6 ± 3.9 mmHg for MAP. Bland-Altman analysis of MAP values showed a bias of 0.26 mmHg, with upper and lower limits of agreement of 12.18 mmHg and -11.67 mmHg, respectively. CNAP readings deviated from arterial values by ≤ 5 mmHg in 67% of MAP values, 59% of DBP values, and 43% of SBP readings. The difference was ≤ 10 mmHg for 94% of MAP readings, 90% of DBP values, and 73% of SBP readings. Conclusions During prone positioning, the CNAP monitor provided clinically acceptable accuracy for MAP values, similar to those reported in adults in the supine position.
机译:研究目的旨在评估俯卧在手术室中的患者的无创连续动脉压(CNAP)监护仪的准确性。设计前瞻性研究。在儿童医院设置手术室。患者20名儿科患者,年龄13.8±2岁,体重63.7±18.8 kg,计划在俯卧位进行手术,并已计划将其放置于动脉导管中。干预措施用动脉管(AL)和新型无创连续血压(BP)监护仪记录测量结果。在麻醉期间,每分钟从动脉套管和CNAP装置获取收缩压(SBP),舒张压(DBP)和平均动脉压(MAP)读数。主要结果该研究队列包括对4104对SBP,DBP和MAP值的分析,显示AL和CNAP设备读数之间的绝对差值分别为:SBP 7.9±6.3 mmHg,DBP 5.3±4.3 mmHg和4.6± MAP的3.9毫米汞柱。 MAP值的Bland-Altman分析显示偏差为0.26 mmHg,一致的上限和下限分别为12.18 mmHg和-11.67 mmHg。 CNAP读数在67%的MAP值,59%的DBP值和43%的SBP读数中偏离动脉值≤5 mmHg。对于94%的MAP读数,90%的DBP值和73%的SBP读数,差异为≤10 mmHg。结论在俯卧位期间,CNAP监护仪可提供MAP值的临床可接受的准确性,类似于在仰卧位成人中报告的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号