PURPOSE: To determine the reproducibility of intraocular lens (IOL) decentration and tilt measurements with a new Purkinje meter instrument. SETTING: Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom. METHODS: After pupil dilation, images of pseudophakic eyes with a plate-style IOL (Akreos Adapt) were obtained using a recently developed Purkinje meter. Intraocular lens decentration and tilt were evaluated by analyzing the captured images using a semiobjective method by marking the reflexes in the images and automatic calculation using a dedicated software program. In study 1, examiner 1 examined the eyes first followed by examiner 2. Ten minutes later, examiner 1 performed a second measurement, after which the intraexaminer and interexaminer reproducibility were determined. In study 2, a Purkinje meter was used to measure pseudophakic eyes with slitlamp finding of clinical IOL decentration, IOL tilt, or both. The results were compared with retroillumination photographs and slitlamp findings. RESULTS: In study 1, there was high intraexaminer reproducibility for decentration (r = 0.95) and tilt (r = 0.85) and high interexaminer reproducibility for decentration (r = 0.84) and tilt (r = 0.75). In study 2, even in extreme cases of decentration and/or tilt, the Purkinje meter measurements were possible and appeared to correlate well with slitlamp findings. CONCLUSIONS: Acquisition of images in pseudophakic eyes with the Purkinje meter was simple and rapid. The method was highly reliable for 1 examiner and between 2 examiners.
机译:目的:使用新型浦肯野仪表来确定人工晶状体(IOL)偏心和倾斜测量的可重复性。地点:英国伦敦Moorfields眼科医院NHS基金会信托基金。方法:瞳孔散大后,使用最新开发的Purkinje测量仪,用板状IOL(Akreos Adapt)获得假晶状体眼的图像。眼内透镜的偏斜和倾斜是通过使用半客观方法分析捕获的图像来评估的,方法是标记图像中的反射并使用专用软件程序自动计算。在研究1中,检查者1首先检查眼睛,然后检查者2。十分钟后,检查者1进行了第二次测量,此后确定了内部检查者和内部检查者的可重复性。在研究2中,使用Purkinje仪通过裂隙灯测量假晶状体眼,以发现临床IOL偏低,IOL倾斜或两者兼有。将结果与逆光照片和裂隙灯结果进行比较。结果:在研究1中,检偏器内的重现性高(r = 0.95)和倾斜度(r = 0.85),检偏器内的重现性高(r = 0.84)和倾斜度(r = 0.75)。在研究2中,即使在极端偏心和/或倾斜的情况下,浦肯野仪的测量也是可能的,并且似乎与裂隙灯的发现密切相关。结论:用浦肯野仪获得假晶状体眼中的图像简单,快速。该方法对于1位检查员和2位检查员之间的可靠性很高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号