Analytical performance evaluation of the scanning capillary tube viscometer for measurement of whole blood viscosity

机译:扫描毛细管粘度计用于全血粘度测量的分析性能评估

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Whole blood viscosity (WBV) is the resistance of blood flow in blood vessels. Increased WBV may be a cardiovascular risk factor. The proper screening of WBV can help the early detection of cardiovascular disease. We investigated the performance of a new scanning capillary tube viscometer (SCTV) for the measurement of WBV. Methods: We evaluated the total precision of the SCTV for 20. days using three control viscosity materials, and the within-day precision with the whole blood samples of three different individuals. For the linearity evaluation, serial dilutions of a high concentration standard material were used. For the method comparison, the results of the SCTV method were compared to those of Brookfield rotating viscometer on 227 subjects. Results: The SCTV had good within-run and total-run coefficient of variant (CV)s at low-, medium-, and high-concentration samples, at shear rates of 1 and 300s-1. The within-day CVs with the three human blood samples were 6.3%, 3.7% and 3.8% at a shear rate of 1s-1, and 3.2%, 3.0% and 4.1% at a shear rate of 300s-1. The SCTV method showed an excellent linearity in the range of 84.9 to 558.2 milliPoise (mP) and 28.8 to 71.0 mP at shear rates of 1 and 300s-1, respectively. For the comparison study, the SCTV and the rotating viscometer showed comparable results. Conclusions: The SCTV showed a stable analytical performance, and was comparable with the rotational viscometer. This new SCTV method can be used in the clinical laboratory for various needs.
机译:背景:全血粘度(WBV)是血管中血流的阻力。 WBV升高可能是心血管危险因素。正确筛查WBV有助于早期发现心血管疾病。我们调查了新型WBV扫描毛细管粘度计(SCTV)的性能。方法:我们使用三种控制粘度材料评估了SCTV在20天之内的总精度,以及三个不同个体的全血样本的日内精度。为了进行线性评估,使用了高浓度标准物质的系列稀释液。为了进行方法比较,对227位受试者的SCTV方法的结果与Brookfield旋转粘度计的结果进行了比较。结果:SCTV在低,中和高浓度样品中,剪切速率分别为1和300s-1时,具有良好的变异内(CV)的运行内和总运行系数。三种人体血液样本的日内CV在1s-1的剪切速率下分别为6.3%,3.7%和3.8%,在300s-1的剪切速率下分别为3.2%,3.0%和4.1%。 SCTV方法在剪切速率分别为1和300s-1时分别在84.9至558.2毫泊(mP)和28.8至71.0 mP的范围内显示出出色的线性。为了进行比较研究,SCTV和旋转粘度计显示了可比较的结果。结论:SCTV显示稳定的分析性能,并且与旋转粘度计相当。这种新的SCTV方法可以在临床实验室中用于各种需求。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号