Corrosion testing of stents: a novel fixture to hold entire device in deployed form and finish.

机译:支架的腐蚀测试:一种新颖的固定装置,可将整个设备固定在展开状态并完成。

获取原文
获取原文并翻译 | 示例
           

摘要

Stainless steel and nickel-titanium alloy stents for cardiac, biliary, and prostatic clinical applications may be susceptible to loss of mechanical integrity/failure due to their small size and unique geometry. Because their corrosion behavior is a critical aspect of their biocompatibility, it is necessary to test such devices for corrosion behavior using a final deployed form and finish condition. Welding wire leads or holding with metallic clips can result in surface variability or damage to the stents. A novel fixture to hold stents for corrosion testing was designed and evaluated in this study. This fixture design minimally influenced the properties of stents, and also facilitated easy removal post-testing for other types of characterization. Copyright 1999 John Wiley & Sons, Inc.
机译:用于心脏,胆道和前列腺临床应用的不锈钢和镍钛合金支架可能因尺寸小和独特的几何形状而容易丧失机械完整性/故障。由于它们的腐蚀行为是其生物相容性的关键方面,因此有必要使用最终展开的形状和完成条件来测试此类设备的腐蚀行为。焊接导线或用金属夹固定可能会导致表面变化或损坏支架。在这项研究中,设计并评估了一种新型支架来固定支架进行腐蚀测试。这种固定装置的设计对支架性能的影响最小,并且还便于在测试后轻松移除其他类型的特征。版权所有1999 John Wiley&Sons,Inc.