首页> 外文期刊>Journal of Biomechanics >Mechanical testing of internal fixation devices: A theoretical and practical examination of current methods
【24h】

Mechanical testing of internal fixation devices: A theoretical and practical examination of current methods

机译:内固定装置的机械测试:当前方法的理论和实践检验

获取原文
获取原文并翻译 | 示例
           

摘要

Successful healing of long bone fractures is dependent on the mechanical environment created within the fracture, which in turn is dependent on the fixation strategy. Recent literature reports have suggested that locked plating devices are too stiff to reliably promote healing. However, in vitro testing of these devices has been inconsistent in both method of constraint and reported outcomes, making comparisons between studies and the assessment of construct stiffness problematic. Each of the methods previously used in the literature were assessed for their effect on the bending of the sample and concordant stiffness. The choice of outcome measures used in in vitro fracture studies was also assessed. Mechanical testing was conducted on seven hole locked plated constructs in each method for comparison. Based on the assessment of each method the use of spherical bearings, ball joints or similar is suggested at both ends of the sample. The use of near and far cortex movement was found to be more comprehensive and more accurate than traditional centrally calculated interfragmentary movement values; stiffness was found to be highly susceptible to the accuracy of deformation measurements and constraint method, and should only be used as a within study comparison method. The reported stiffness values of locked plate constructs from in vitro mechanical testing is highly susceptible to testing constraints and output measures, with many standard techniques overestimating the stiffness of the construct. This raises the need for further investigation into the actual mechanical behaviour within the fracture gap of these devices. (C) 2015 Elsevier Ltd. All rights reserved.
机译:长骨骨折的成功治愈取决于骨折内产生的机械环境,而机械环境又取决于固定策略。最近的文献报道表明,锁定的电镀装置太硬而不能可靠地促进愈合。但是,这些装置的体外测试在约束方法和已报道的结果方面均不一致,这使得研究与构造刚度评估之间的比较存在问题。评估先前文献中使用的每种方法对样品弯曲和一致硬度的影响。还评估了体外骨折研究中使用的结局指标的选择。为了进行比较,在每种方法中对七个孔锁定的电镀结构进行了机械测试。根据对每种方法的评估,建议在样品的两端使用球形轴承,球形接头或类似零件。人们发现,使用远近皮层运动比传统的中央计算的碎片间运动值更全面,更准确。刚度对变形测量和约束方法的准确性非常敏感,应仅用作研究比较方法。通过体外机械测试报告的锁板结构的刚度值极易受到测试约束和输出量度的影响,许多标准技术都高估了该结构的刚度。这就需要进一步研究这些装置的断裂间隙内的实际机械性能。 (C)2015 Elsevier Ltd.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号