Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure.

机译:脉搏到达时间不足以替代脉搏传播时间作为血压的指标。

获取原文
获取原文并翻译 | 示例
           

摘要

Pulse transit time (PTT) is a proven, simple to measure, marker of blood pressure (BP) that could potentially permit continuous, noninvasive, and cuff-less BP monitoring (after an initial calibration). However, pulse arrival time (PAT), which is equal to the sum of PTT and the pre-ejection period, is gaining popularity for BP tracking, because it is even simpler to measure. The aim of this study was to evaluate the hypothesis that PAT is an adequate surrogate for PTT as a marker of BP. PAT and PTT were estimated through the aorta using high-fidelity invasive arterial waveforms obtained from six dogs during wide BP changes induced by multiple interventions. These time delays and their reciprocals were evaluated in terms of their ability to predict diastolic, mean, and systolic BP (DBP, MBP, and SBP) per animal. The root mean squared error (RMSE) between the BP parameter predicted via the time delay and the measured BP parameter was specifically used as the evaluation metric. Taking the reciprocals of the time delays tended to reduce the RMSE values. The DBP, MBP, and SBP RMSE values for 1/PAT were 9.8 +/- 5.2, 10.4 +/- 5.6, and 11.9 +/- 6.1 mmHg, whereas the corresponding values for 1/PTT were 5.3 +/- 1.2, 4.8 +/- 1.0, and 7.5 +/- 2.2 mmHg (P < 0.05). Thus tracking BP via PAT was not only markedly worse than via PTT but also unable to meet the FDA BP error limits. In contrast to previous studies, our results quantitatively indicate that PAT is not an adequate surrogate for PTT in terms of detecting challenging BP changes.
机译:脉搏传播时间(PTT)是一种经过验证的,易于测量的血压标记(BP),可以潜在地允许连续,无创且无袖带的BP监测(在初始校准之后)。但是,等于PTT和预喷射时间之和的脉冲到达时间(PAT)在BP跟踪中越来越受欢迎,因为它甚至更易于测量。这项研究的目的是评估以下假设:PAT是作为BP标记物的PTT的适当替代品。通过使用六种狗在多次干预引起的广泛BP变化过程中获得的高保真侵入性动脉波形,通过主动脉估计PAT和PTT。根据它们预测每只动物的舒张压,均值和收缩压(DBP,MBP和SBP)的能力,评估了这些时间延迟及其倒数。通过时间延迟预测的BP参数与测得的BP参数之间的均方根误差(RMSE)特别用作评估指标。采取时间倒数的倒数往往会降低RMSE值。 1 / PAT的DBP,MBP和SBP RMSE值为9.8 +/- 5.2、10.4 +/- 5.6和11.9 +/- 6.1 mmHg,而1 / PTT的相应值为5.3 +/- 1.2、4.8 +/- 1.0和7.5 +/- 2.2 mmHg(P <0.05)。因此,通过PAT跟踪BP不仅明显比通过PTT跟踪差,而且无法满足FDA BP错误限制。与以前的研究相比,我们的结果定量地表明,就检测具有挑战性的BP变化而言,PAT并不是PTT的适当替代品。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号