A triiodide-selective electrode based on copper (II)-Schiff base complex as a membrane carrier is proposed. The electrode was prepared by incorporating the carrier into a plasticized polyvinylchloride (PVC) membrane and was directly coated on the surface of a graphite electrode. The obtained electrode showed a near Nernstian slope of 57.0 +/- 0.4 mV/decade to I-3(-) ions over an activity range of 1.0 x 10(-5)-1.0 x 10(-1) M with a limit of detection of 4.8 x 10(-6) M. The response time of the electrode was fast (5 s) and the electrode could be used for about 2 months without considerable divergence in response. The potentiometric selectivity coefficients were evaluated and displayed anti-Hofmeister behavior. The electrode was used as an indicator electrode in the potentiometric titration of the triiodide ion and in the determination of ascorbic acid in vitamin C tablets.
机译:提出了一种基于铜-席夫碱配合物作为膜载体的三碘化物选择性电极。通过将载体掺入增塑的聚氯乙烯(PVC)膜中来制备电极,并将其直接涂覆在石墨电极的表面上。所获得的电极在1.0 x 10(-5)-1.0 x 10(-1)M的活动范围内,对I-3(-)离子的Nernstian斜率接近57.0 +/- 0.4 mV /十倍。检测到4.8 x 10(-6)M。该电极的响应时间很快(5 s),该电极可以使用约2个月,而响应差异不大。评估电位选择系数并显示出抗霍夫迈斯特行为。该电极在三碘化物离子的电位滴定和维生素C片剂中抗坏血酸的测定中用作指示电极。