A microperimeter that provides fixation pattern and retinal sensitivity measurement.

机译:提供固定模式和视网膜灵敏度测量的微周边。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To investigate a microperimeter (MP-1) that can independently measure fixation status and microperimetry of retinal threshold sensitivity by comparing the results obtained by the MP-1 with those obtained by scotometry using scanning laser ophthalmoscopy (SLO). METHODS: Fourteen patients (15 eyes) with a macular disorder were examined by the MP-1 fixation test, MP-1 microperimetry, and SLO scotometry. The scotoma size seen in MP-1 microperimetry was compared with that observed in SLO scotometry. The location of the preferred retinal locus (PRL) and the fixation stability, which were obtained from the fixation test in MP-1, were also compared with those obtained in SLO scotometry. RESULTS: The scotoma size, which was not seen with 0 dB in MP-1 microperimetry, was larger than that in SLO scotometry in 8 of the 15 eyes. Retinal threshold sensitivity decreases were found using MP-1 microperimetry within a sensitive area of SLO scotometry in all eyes. The PRL in the MP-1 fixation test and in SLO scotometry agreed in all eyes. Fixation stability in the MP-1 fixation test significantly correlated with that in SLO scotometry (P 0.0192). CONCLUSIONS: The results of MP-1 microperimetry do not completely agree with those of SLO scotometry owing to the difference in stimulus intensity. The MP-1 fixation test is useful for measuring PRL and fixation stability in a short time. MP-1 might be helpful to evaluate the foveal function in patients with macular disorders without severely damaged macular function.
机译:目的:通过比较MP-1获得的结果与使用扫描激光检眼镜(SLO)进行的透视法获得的结果,研究可独立测量视网膜状态阈值敏感性的固定状态和显微视野的微视野(MP-1)。方法:通过MP-1固定测试,MP-1显微视野测定法和SLO苏格兰比色法对14例(15眼)黄斑疾病患者进行了检查。将MP-1显微透视法中观察到的暗影大小与SLO立体法中观察到的暗影大小进行了比较。还将从MP-1中的固定测试获得的首选视网膜位点(PRL)的位置和固定稳定性与通过SLO立体法获得的视网膜位置进行了比较。结果:在15只眼中的8只眼中,在MP-1显微视野法中0 dB看不到的暗点大小比SLO暗度法大。在所有眼睛的SLO眼压测定法的敏感区域内使用MP-1显微视野测定法发现视网膜阈值敏感性降低。 MP-1固定测试和SLO苏格兰比光法中的PRL在所有眼睛中都一致。 MP-1固定测试中的固定稳定性与SLO苏格兰比色法中的固定稳定性显着相关(P 0.0192)。结论:由于刺激强度的差异,MP-1显微视野测定的结果与SLO容积测定的结果并不完全一致。 MP-1固定测试可用于在短时间内测量PRL和固定稳定性。 MP-1可能有助于评估没有严重损害黄斑功能的黄斑疾病患者的中央凹功能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号