Noninvasive LV pressure estimation using subharmonic emissions from microbubbles

机译:利用微气泡的亚谐波发射进行无创低压压力估计

获取原文
获取原文并翻译 | 示例
           

摘要

To develop a new noninvasive approach to quantify left ventricular (LV) pressures using subharmonic emissions from microbubbles, an ultrasound scanner was used in pulse inversion grayscale mode; unprocessed radiofrequency data were obtained with pulsed wave Doppler from the aorta and/or LV during Sonazoid infusion. Subharmonic data (in dB) were extracted and processed. Calibration factor (mm Hg/dB) from the aortic pressure was used to estimate LV pressures. Errors ranged from 0.19 to 2.50 mm Hg when estimating pressures using the aortic calibration factor, and were higher (0.64 to 8.98 mm Hg) using a mean aortic calibration factor. Subharmonic emissions from ultrasound contrast agents have the potential to noninvasively monitor LV pressures.
机译:为了开发一种新的非侵入性方法来利用微气泡的次谐波发射来量化左心室(LV)压力,在脉冲反转灰度模式下使用了超声扫描仪。在Sonazoid输注过程中,使用脉冲波多普勒从主动脉和/或LV获得未处理的射频数据。提取并处理次谐波数据(以dB为单位)。主动脉压力的校准因子(mm Hg / dB)被用来估计左室压力。使用主动脉校准因子估算压力时,误差范围为0.19至2.50 mm Hg,使用平均主动脉校准因子估算时,误差较高(0.64至8.98 mm Hg)。超声造影剂的亚谐波发射具有无创监测左室压力的潜力。