首页> 外文期刊>The Journal of arthroplasty >Post damage in contemporary posterior-stabilized tibial inserts: influence of implant design and clinical relevance.
【24h】

Post damage in contemporary posterior-stabilized tibial inserts: influence of implant design and clinical relevance.

机译:当代后稳定胫骨插入物的后损伤:植入物设计和临床相关性的影响。

获取原文
获取原文并翻译 | 示例
           

摘要

The mechanisms of damage at the polyethylene post in 3 contemporary tibial insert designs were evaluated and compared with a historical standard (Insall-Burstein II; Zimmer, Warsaw, Ind). One hundred five gamma sterilized posterior-stabilized tibial inserts were revised after an average of 4.7 years (0.05-13.6 years). Retrievals were classified according to their designs: Insall-Burstein II (n = 28); PFC (Johnson & Johnson, Raynham, Mass; n = 30); NexGen (Zimmer; n = 32); and Scorpio (Stryker Orthopaedics, Mahwah, NJ; n = 15). Reasons for revision and patient details were available. Surface damage scoring and photogrammetry were performed on all the retrieved tibial inserts. Oxidation analysis was carried out for traceable historical, gamma air-sterilized and conventional, gamma inert-sterilized tibial inserts (n = 61) with the use of infrared spectroscopy. The posts for all 3 contemporary designs exhibited damage similar to the historical controls. Articular, post, and backside damage scores significantly increased with implantation time. Post damage was insensitive to design and patient factors but was exacerbated by oxidation. An association between damage at the post and articular surface was also confirmed. Logistic models suggested an interaction between post damage, backside surface damage, and implant loosening.
机译:评估了3种当代胫骨插入物设计中聚乙烯桩的损伤机理,并将其与历史标准(Insall-Burstein II; Zimmer,华沙,印第安纳州)进行了比较。平均4.7年(0.05-13.6年)后,对105例经伽玛消毒的后稳定胫骨插入物进行了修订。取物根据其设计分类:Insall-Burstein II(n = 28); PFC(马萨诸塞州雷纳姆市的强生公司; n = 30); NexGen(Zimmer; n = 32);和天蝎座(Stryker Orthopaedics,新泽西州Mahwah; n = 15)。可以提供修订原因和患者详细信息。对所有取回的胫骨插入物进行表面损伤评分和摄影测量。使用红外光谱法对可追溯的历史,伽玛空气灭菌和常规伽玛惰性灭菌的胫骨插件(n = 61)进行了氧化分析。所有3种现代设计的帖子都显示出与历史控件类似的损坏。关节,后和后侧的损伤评分随植入时间的延长而显着增加。损伤后对设计和患者因素不敏感,但由于氧化而加剧。还证实了在柱的损伤和关节表面之间的关联。 Logistic模型表明桩后破坏,背面表面破坏和植入物松动之间存在相互作用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号