首页> 外文期刊>The Journal of arthroplasty >Five Duraloc locking ring failures.
【24h】

Five Duraloc locking ring failures.

机译:五次Duraloc锁定环故障。

获取原文
获取原文并翻译 | 示例
           

摘要

We report on 5 cases that underwent revision for locking ring failure in the Duraloc product line (DePuy, Warsaw, Ind). All liner retrievals showed signs of posterior neck/liner impingement and superior edge loading or significant wear. In these cases, we believe superior head migration and neck/liner impingement due to cup anteversion contributed to these locking ring failures. More research is needed to determine the incidence of this complication. Patients with locking ring failures should be closely monitored. Full cup revision, face-changing liners, or cementing liners into well-fixed cups are all options to correct suboptimal cup positioning.
机译:我们报告了5例因Duraloc产品线(DePuy,Warsaw,Ind)的锁紧环失效而进行了修订的案例。所有的内衬取回均显示后颈部/内衬撞击的迹象,边缘负荷较高或明显磨损。在这些情况下,我们认为由于杯前倾而导致的头部移行和颈部/内胆撞击效果强是造成这些锁紧环失效的原因。需要更多的研究来确定这种并发症的发生率。锁紧环失败的患者应受到密切监测。全杯翻新,换脸衬套或固定在固定杯中的固结衬套都是纠正次优杯定位的选择。