Catastrophic failure of a modular revision total hip polyethylene insert.

机译:模块化改型全髋部聚乙烯插件发生灾难性故障。

获取原文
获取原文并翻译 | 示例
           

摘要

Early catastrophic failure of a modular polyethylene component is a potential problem after revision total hip arthroplasty. We describe an unusual case of polyethylene failure that occurred within 18 months of implantation in which no obvious technical error or mechanical failure was identified. The acetabular polyethylene insert was prepared with gas plasma sterilization, and the shelf life was 4 months. Radiographic evaluation used generalized Hough transforms to assess the cup articulation. We identified cup penetration of 2.8 mm before revision and catastrophic destruction of the polyethylene liner at the time of revision. Possible factors implicated for failure included a thin polyethylene liner, increased hip separation, femoral head mismatch, and the high activity level of a younger patient. We believe that this case report highlights the need for future investigations of these subtle factors.
机译:翻修全髋关节置换术后,模块化聚乙烯组件的早期灾难性故障是一个潜在的问题。我们描述了在植入后18个月内发生的聚乙烯失效的异常情况,其中未发现明显的技术错误或机械故障。髋臼聚乙烯插入物通过气体等离子体灭菌制备,保质期为4个月。射线照相评估使用广义霍夫变换来评估杯的清晰度。我们发现翻新前的杯穿透度为2.8 mm,翻新时聚乙烯衬里发生了灾难性破坏。可能导致失败的因素包括聚乙烯衬里薄,髋部分离增加,股骨头错配以及年轻患者的高活动水平。我们认为,本案例报告强调了对这些微妙因素进行进一步调查的必要性。