Effect of cementless acetabular component orientation, position, and containment in total hip arthroplasty for congenital hip disease.

机译:非骨水泥髋臼组件的方向,位置和封闭在先天性髋关节疾病全髋关节置换术中的作用。

获取原文
获取原文并翻译 | 示例
           

摘要

We evaluated the effect of the inclination angle, position, and containment of 53 cementless cups inserted in patients with congenital hip disease (CHD), after a minimum of 10 years of follow-up. The polyethylene wear rate was significantly greater when the cup was placed in more than 45 degrees inclination (P=.045) or if the cup was placed lateral to the teardrop position by more than 25 mm (P=.001). Aseptic loosening of the femoral component was significantly greater when the cup was placed more than 25 mm superiorly to the teardrop (P=.049). Cup placement of more than 25 mm lateral to the teardrop affected significantly periacetabular osteolysis (P=.032). In CHD cases, it is preferable to avoid excessive vertical inclination, lateral, and superior placement of cementless cups in an attempt to obtain better containment.
机译:经过至少10年的随访,我们评估了先天性髋部疾病(CHD)患者中插入的53个非骨水泥杯的倾斜角度,位置和围堵的效果。当将杯子倾斜放置超过45度时(P = .045)或如果将杯子放置在水滴位置的侧面超过25 mm(P = .001),则聚乙烯的磨损率会大大提高。当将杯子放置在距泪珠上方25毫米以上的位置时,股骨组件的无菌性松动明显更大(P = .049)。泪滴侧面25毫米以上的杯子放置会明显影响髋臼周围骨溶解(P = .032)。在冠心病的情况下,最好避免过度的垂直倾斜,侧面和非水泥杯的过度放置,以期获得更好的密封性。