首页> 外文期刊>The Journal of arthroplasty >A study of polyethylene and modularity issues in >1,000 posterior cruciate-retaining knees at 5 to 11 years.
【24h】

A study of polyethylene and modularity issues in >1,000 posterior cruciate-retaining knees at 5 to 11 years.

机译:在5至11岁时对超过1,000个后交叉保留膝盖的聚乙烯和模块化问题进行了研究。

获取原文
获取原文并翻译 | 示例
           

摘要

Polyethylene quality and production technique and monoblock versus modular design of tibial components affect the polyethylene wear of total knee arthroplasties (TKAs). We reviewed >1,000 TKAs performed with the AGC prosthesis (Biomet, Warsaw, IN) using a uniform surgical technique with a 5- to 11-year follow-up. Of tibial components, 698 were of compression-molded monoblock design and 353 were of ram-extruded modular assembly design. The effect of the differences in tibial component design on the 5- to 11-year clinical and radiologic outcomes was studied. Higher rates of osteolysis, radiolucent lines, and revision were found with the ram-extruded modular design. This modular design is similar to most tibial component designs currently used in TKA, yet osteolysis in completely cemented components is described rarely. The TKAs done with a compression-molded monoblock design gave better clinical and radiologic results. The improved results may be due to polyethylene quality, assembly design, or a combination of these factors.
机译:聚乙烯的质量和生产技术以及胫骨组件的整体式与模块化设计会影响全膝关节置换术(TKA)的聚乙烯磨损。我们回顾了使用AGC假体(Biomet,华沙,印第安纳州)使用统一的手术技术进行了超过1,000例TKA,并进行了5至11年的随访。在胫骨组件中,有698件采用压模整体成型设计,而353件采用夯锤挤压式模块化装配设计。研究了胫骨组件设计差异对5至11年临床和影像学结果的影响。冲压挤压的模块化设计发现了更高的溶骨率,射线可透性线和翻修率。这种模块化设计与目前在TKA中使用的大多数胫骨组件设计相似,但很少描述完全粘合组件中的骨溶解。用压缩模制整体设计完成的TKA提供了更好的临床和放射学结果。改进的结果可能归因于聚乙烯质量,组件设计或这些因素的组合。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号