摘要

We performed 204 cementless total hip arthroplasties using a fully porous stem made of a cast cobalt-chrome-molybdenum alloy. Five stems fractured at the middle part. Champagne-fluted canals (P<.0001) and low canal fillings 1 cm below the lesser trochanter (P =.02) significantly correlated with stem fractures. Subsequent surgery revealed that all of the proximal parts were surrounded by fibrous tissue, and the distal parts showed bone ingrowth. Numerous voids were present close to the surface of the implant body. The core diameters of the fractured stems were 4 to 5 mm. The fractures may be attributed to the combination of the lack of proximal support, a champagne-fluted canal, the fully porous stem made of a cast cobalt-chrome-molybdenum alloy, and the narrow dimension of the stem core.
机译:我们使用由铸造钴铬钼合金制成的完全多孔的茎杆进行了204次非骨水泥全髋关节置换术。中部有五个茎断裂。香槟槽(P <.0001)和小转子下方1 cm处的低运河充盈量(P = .02)与茎干骨折显着相关。随后的手术显示,所有近端部分均被纤维组织包围,而远端部分则显示出骨向内生长。植入物表面附近存在大量空隙。断裂茎的核心直径为4至5毫米。骨折的原因可能是缺乏近端支撑,没有香槟槽的管,由铸造钴铬钼合金制成的完全多孔的杆以及杆芯的狭窄尺寸。