Anteversion and vertical tilt of the acetabular prostheses in 50 consecutive total hip arthroplasties were prospectively evaluated during surgery (by the surgeon, using an alignment guide) and radiographically (calculated). From postoperative standardized radiographs vertical tilt was measured directly and anteversion was calculated. The mean error of vertical tilt was 5 degrees (range, 0 degrees - 20 degrees). The mean error of version was 9 degrees (range, 0 degrees - 24 degrees). The reliability of prosthesis placement in a predetermined zone was examined. Although the surgeons believed that all 50 cups were inside this zone, radiographic measurements revealed that 21 of the cups were actually outside. It is concluded that vertical tilt can be reasonably assessed during surgery. Anteversion, however, cannot be accurately assessed during surgery, despite use of the alignment guide.
机译:前瞻性评估手术期间(由外科医生使用对准导引器)进行的50次连续全髋关节置换术中髋臼假体的前倾角和垂直倾斜度,并进行射线照相(计算)。从术后标准化X射线照片直接测量垂直倾斜并计算前倾角。垂直倾斜的平均误差为5度(范围为0度至20度)。版本的平均误差为9度(范围为0度至24度)。检查假体在预定区域内放置的可靠性。尽管外科医生认为所有50个杯子都在该区域内,但射线照相测量显示其中21个杯子实际上在外面。结论是可以在手术期间合理地评估垂直倾斜。但是,尽管使用了对齐指南,但在手术过程中无法准确评估前屈。