首页> 外文期刊>The Journal of heart valve disease >Digital frequency analysis of valve sound phenomena in patients after prosthetic valve surgery: its capability as a true home monitoring of valve function.
【24h】

Digital frequency analysis of valve sound phenomena in patients after prosthetic valve surgery: its capability as a true home monitoring of valve function.

机译:人工瓣膜手术后患者瓣膜声音现象的数字频率分析:其作为真正的家庭瓣膜功能监测的能力。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND AND AIM OF THE STUDY: Depending on the individual risk profile of a patient, disturbances of the functional integrity of mechanical heart valve prostheses occur in up to 2.5% of patients each year. The early phase of prosthetic dysfunction (due to thrombus formation, tissue ingrowth or endocarditis) usually remains undiagnosed, as patients do not present with symptoms in this situation, and imaging techniques (echocardiography, fluoroscopy) demonstrate normal occluder motion. The delay between the onset of prosthetic valve dysfunction and its clinical manifestation may result in complications (e.g. thromboembolism) or extended therapeutic options (e.g. reoperation rather than more intensive anticoagulation). METHODS: A total of 291 patients with mechanical heart valves was allocated to four different sub-groups, and each measured their valve sounds regularly with the 'ThromboCheck' device. Depending on the subgroup, the signals were compared with different reference signals. Patients in whoma suspicious signal was detected were immediately contacted and examined meticulously. RESULTS: Fourteen patients were found to have suspicious signals. In 13 patients, valve dysfunction was confirmed by fluoroscopy, but in four cases neither transthoracic nor transesophageal echocardiography detected abnormal occluder motion or 'musses' adjacent to the prosthesis. Normal valve sounds returned in four patients who underwent thrombolytic therapy. All patients regularly recorded and passed on their signals. Surveys revealed high acceptance and easy handling of the Thrombocheck device. CONCLUSION: Home monitoring of sound pressure measurements of prosthetic valves by digital frequency analysis via a Fast Fourier transformation may detect even very mild alterations of prosthetic valve function. The next evolution of control systems, allowing for registration of flow, frequency spectrum and electrocardiography, opens potential applications for Internet-based, remote monitoring of cardiac patients.
机译:研究背景和目的:根据患者的个人风险状况,每年多达2.5%的患者会发生机械性心脏瓣膜假体功能完整性的紊乱。假体功能障碍的早期阶段(由于血栓形成,组织向内生长或心内膜炎)通常仍未得到诊断,因为患者在这种情况下没有症状,并且成像技术(超声心动图,荧光检查)显示出正常的封堵器运动。人工瓣膜功能障碍发作与其临床表现之间的延迟可能会导致并发症(例如血栓栓塞)或扩展的治疗选择(例如再次手术,而不是更强烈的抗凝治疗)。方法:将总共291例机械性心脏瓣膜患者分配到四个不同的子组,每个小组均使用“ ThromboCheck”设备定期测量其瓣膜声音。根据子组,将信号与不同的参考信号进行比较。立即检查发现有可疑信号的患者,并进行仔细检查。结果:发现十四名患者有可疑信号。在13例患者中,通过透视检查证实了瓣膜功能障碍,但在4例中,经胸或经食道超声心动图均未检测到邻近假体的封堵器运动异常或“肌肉”。接受溶栓治疗的四名患者的瓣膜声音恢复正常。所有患者定期记录并传递信号。调查显示Thrombocheck设备的接受程度很高且易于操作。结论:通过快速傅立叶变换通过数字频率分析对人工瓣膜的声压测量进行家庭监测,甚至可以检测到人工瓣膜功能的非常轻微的变化。控制系统的下一次发展,可以记录血流,频谱和心电图,为基于Internet的心脏病患者远程监控打开了潜在的应用领域。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号