Tissue response to surgical energy devices.

机译:组织对手术能量装置的反应。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: There is little data available on the effects of energy-based surgical devices (ESD) on tissues other than arteries and veins. As such, we quantified the lateral thermal damage associated with contemporary ESD: the Harmonic ACE, a prototype bipolar device known as the Gyrus Trisector, the Harmonic LCS-C5, and the LigaSure V. METHODS: We divided 24 domestic pigs into 4 groups, 1 group for each ESD tested. Segments of bladder, stomach, small bowel, colon, ureter, peritoneum, arteries, and veins were exposed to each ESD. The tissues were stained with hematoxylin and eosin and evaluated by an experienced pathologist to quantitate the lateral energy spread associated with each device. We measured blade temperatures of each device using the IR-Flex thermal camera. RESULTS: The Trisector developed the lowest mean blade temperature (97.84 degrees F), whereas the LigaSure's was the second lowest (103.14 degrees F). The ACE and LCS-C5 created the highest blade temperatures, measuring 220.5 degrees F and 205.6 degrees F, respectively. The Trisector's mean full thickness and superficial lateral energy damage were 6.3 mm and 7.0 mm, respectively, whereas the Ligasure's was 4.5 mm and 5.9 mm, respectively. For the ACE, however, mean full thickness and superficial energy spread were 2.4 mm and 2.8 mm, respectively, whereas the LCS-C5's were 3.1 mm and 4.3 mm, respectively. CONCLUSIONS: The Harmonic ACE and LCS-C5 produced the least thermal damage in the tissues tested. ESD-associated tissue energy damage is not directly related to blade temperature, but is likely the result of several factors including blade temperature, transection time, tissue properties, and the vascularity of each transected tissue.
机译:目的:关于基于能量的外科手术设备(ESD)对动脉和静脉以外的组织的影响的数据很少。因此,我们量化了与当代ESD相关的侧面热损害:Harmonic ACE,一种称为Gyrus Trisector的双极原型设备,Harmonic LCS-C5和LigaSureV。方法:我们将24只家猪分为4组,每个ESD测试一组。每种ESD均暴露了膀胱,胃,小肠,结肠,输尿管,腹膜,动脉和静脉的部分。用苏木精和曙红对组织进行染色,并由经验丰富的病理学家进行评估,以定量与每个设备相关的横向能量分布。我们使用IR-Flex热像仪测量了每个设备的刀片温度。结果:Trisector产生了最低的平均刀片温度(97.84华氏度),而LigaSure的刀片平均温度第二低(103.14华氏度)。 ACE和LCS-C5产生了最高的刀片温度,分别为220.5华氏度和205.6华氏度。 Trisector的平均全厚度和表面侧向能量损伤分别为6.3 mm和7.0 mm,而Ligasure的分别为4.5 mm和5.9 mm。但是,对于ACE,平均全厚度和表面能量散布分别为2.4 mm和2.8 mm,而LCS-C5的平均总厚度和表面能量散布分别为3.1 mm和4.3 mm。结论:谐波ACE和LCS-C5在被测组织中产生的热损伤最小。 ESD相关的组织能量损害与刀片温度没有直接关系,而可能是多种因素的结果,包括刀片温度,横切时间,组织特性和每个横切组织的血管性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号