首页> 外文期刊>Proceedings of the Institution of Mechanical Engineers, Part H. Journal of Engineering in Medicine >Spectral analysis of the sound produced during femoral broaching and implant insertion in uncemented total hip arthroplasty
【24h】

Spectral analysis of the sound produced during femoral broaching and implant insertion in uncemented total hip arthroplasty

机译:非骨水泥型全髋关节置换术中股骨钻孔和植入物插入过程中产生的声音的频谱分析

获取原文
获取原文并翻译 | 示例
           

摘要

Preparation of the proximal femur using incremental broaches to create the ideal cancellous bone envelope is an important technique to perfect in uncemented hip arthroplasty. To guide broaching adequacy and final implant position, the surgeon can use audible pitch changes produced by the femoral broach and definitive implant. The aim of this study was to characterise these pitch changes by analysing the sound spectra created by the first broach, last broach and implant using spectral analysis software. The last broach and implant introduction spectra demonstrated low-frequency (400- 1200 Hz) spectral peaks that were not detected when using the first broach. These frequencies corresponded to the natural resonant frequency of a standing sound wave within the femoral bone canal (~894 Hz) that was estimated using acoustic physics theory. The remaining spectral peaks were associated with transverse vibration modes produced by striking the metal broach handle and implant introducer and were a function of the constructs geometry and material properties.
机译:使用增量式开胸器创建理想的松质骨包膜来准备股骨近端是一项重要的技术,可以完美地完成非骨水泥型髋关节置换术。为了指导拉削的充分性和最终植入物的位置,外科医生可以使用股骨拉削和确定的植入物产生的可闻音高变化。这项研究的目的是通过使用频谱分析软件分析第一个拉刀,最后一个拉刀和植入物创建的声谱来表征这些音高变化。最后的拉刀和植入物引入光谱显示出低频(400-1200 Hz)的光谱峰,使用第一个拉刀时未检测到。这些频率对应于股骨管内驻声波的自然共振频率(〜894 Hz),该频率是使用声学物理学原理估算的。其余的光谱峰值与通过敲打金属拉刀手柄和植入物导引器产生的横向振动模式有关,并且是结构几何形状和材料特性的函数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号