首页> 外文期刊>Journal of Hydraulic Engineering >Calibration Accuracy for Constant Temperature Thermal Anemometer
【24h】

Calibration Accuracy for Constant Temperature Thermal Anemometer

机译:恒温热风速计的校准精度

获取原文
获取原文并翻译 | 示例
           

摘要

Thermal anemometry is widely used to measure fluid velocity and boundary shear stress. Sensor calibration is required and results in a nonlinear equation with anemometer voltage as a function of fluid velocity or shear stress. Uncertainties in the parameters of the nonlinear equation are a source of uncertainty in measured values. A method for quantifying the calibration accuracy is presented using regression analysis with 95% confidence and prediction intervals. The proposed method is applied to shear stress measurements of a calibrated flush-mounted hot-film sensor. Shear stresses ranged from 0.6 to 6.8 Pa. Calibration accuracy was found to be within 15% of the absolute shear.
机译:热风速测量法广泛用于测量流体速度和边界切应力。需要进行传感器校准,并生成一个非线性方程,其中风速计电压是流体速度或剪切应力的函数。非线性方程参数的不确定性是测量值不确定性的来源。使用具有95%置信度和预测间隔的回归分析,提出了一种量化校准精度的方法。所提出的方法被应用于校准的齐平安装的热膜传感器的切应力测量。剪切应力范围为0.6到6.8 Pa。发现校准精度在绝对剪切力的15%以内。