首页> 外文期刊>International journal of RF and microwave computer-aided engineering >S-band five-port ring reflectometer-probe system for in vitro breastrntumor detection
【24h】

S-band five-port ring reflectometer-probe system for in vitro breastrntumor detection

机译:用于体外乳腺癌检测的S波段五端口环形反射仪-探头系统

获取原文
获取原文并翻译 | 示例
           

摘要

A five-port ring reflectometer-probe system was developed to measure relative permittivityrnof normal and tumorous breast tissues in frequency range from 1.6 torn2.75 GHz. The calibration procedures of the study reflectometer and the coaxial probernwere described in detail. In this study, the synthetic breast tissues have been used,rnwhich are the Triton TX-100-water mixture at different ratios that mimic dielectricrnproperties of normal and tumor breast tissue samples. The comparative studies ofrnreflection coefficient and dielectric properties of the synthetic mixture samples werernconducted using study reflectometer-probe system and commercial vector networkrnanalyzer with Keysight (formerly Agilent) 85070E dielectric probe. Maximum meanrnerrors of the linear magnitude, phase, dielectric, and loss factor measurements werernrecorded to be 0.02, 6.18, 2.21, and 1.60, respectively. These measurements were significantlyrndistinguishable for normal and tumorous breast tissues.
机译:开发了五端口环形反射仪-探针系统,以测量频率范围为1.6至2.75 GHz的正常和肿瘤性乳腺组织的相对介电常数。详细介绍了研究型反射仪和同轴探头的校准程序。在这项研究中,已使用了合成的乳腺组织,它们是Triton TX-100-水混合物,具有不同的比例,可模仿正常和肿瘤乳腺组织样品的介电特性。使用研究型反射计-探针系统和商用矢量网络分析仪与Keysight(原安捷伦)85070E介电探针对合成混合物样品的反射系数和介电性能进行了比较研究。线性幅度,相位,介电常数和损耗因子测量的最大平均误差分别记录为0.02、6.18、2.21和1.60。这些测量值对于正常和肿瘤性乳腺组织有明显的区别。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号