A simple and expeditious method for the determination of trace metals (cadmium, chromium, copper, manganese and lead) is proposed. The metals are extracted from their matrix by using nitric acid and hydrogen peroxide in a closed-vessel microwave digestion system for their subsequent detection by graphite furnace atomic absorption spectrometry (GFAAS). The sample preparation procedure facilitates the overall analytical process and enables the construction of calibration curves from inorganic standards. The ensuing method provides good linearity and sensitivity for the five metals, with limits of detection and quantization spanning the ranges 0.05-2.20 and 0.15-7.34 μg/kg, respectively. This sensitivity level is quite appropriate for the intended application. Accuracy was assessed by using a certified reference material (NCS ZC85006 Tomato), for which the proposed method provided amounts of metals consistent with their certified values. The proposed method was applied to tomato, pepper and onion, which are widely consumed in Mediterranean countries.
机译:提出了一种简便快速的测定痕量金属(镉,铬,铜,锰和铅)的方法。在封闭容器微波消解系统中,通过使用硝酸和过氧化氢从基质中提取金属,然后通过石墨炉原子吸收光谱法(GFAAS)对其进行检测。样品制备程序有助于整个分析过程,并能够从无机标准品构建校准曲线。随后的方法为五种金属提供了良好的线性和灵敏度,检测和定量的极限分别在0.05-2.20和0.15-7.34μg/ kg的范围内。此灵敏度级别非常适合预期的应用。通过使用认证的参考物质(NCS ZC85006番茄)评估准确性,为此,所建议的方法提供的金属量与其认证值相符。所提出的方法适用于番茄,胡椒和洋葱,这些在地中海国家被广泛食用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号