Ergonomically Flawed Electrode Design Can Create Delays in Defibrillation

机译:符合人体工程学的缺陷电极设计可能会导致除颤延迟

获取原文
           

摘要

Introduction Time is the most important factor affecting survival from ventricular fibrillation cardiac arrest.1 Defibrillation survival decreases by 7-10% for every additional minute of delay.2 Consequently, international guidelines now endorse a goal of only 2-4 minutes from collapse until defibrillation for in-hospital cardiac arrests.2 Cardiac arrest teams are designed to coordinate a timely response, especially to ventricular fibrillation arrest. Unfortunately, excessive delays may occur because the team may be slow to transport, set up, or activate defibrillation equipment. Delays in defibrillation should be viewed for what they are -- a perilous source of medical error. Intuitively one would expect to attribute most defibrillation delays to human error. However, the tremendous impact that flawed equipment design has on medical system error in general is now being increasingly recognized.3 We seek to demonstrate through the following series that flawed defibrillation electrode design can lead directly to intolerable defibrillation delays. Case Reports Our institution uses either Quik-Combo? (Medtronic, Redmond WA) defibrillator electrodes or Padpro? defibrillator electrodes (model 2516 Padpro LLC, Ann Arbor MI) hooked up to a Lifepak? 12 defibrillator (Medtronic, Redmond WA). Unlike the Quik-Combo, the Padpro electrodes require a customized plug adaptor to be connected in series between it and the Lifepak 12 therapy cable (see figure 1). We determined that medical staff confusion over the Padpro and its additional adaptor created unjustifiable delays in defibrillation for two patients.
机译:前言时间是影响心室纤颤心脏骤停生存的最重要因素。1除颤每延误一分钟,除颤生存率就会降低7-10%。2因此,国际准则现在认可从崩溃到除颤只有2-4分钟的目标。 2心脏骤停小组旨在协调及时的响应,尤其是对室颤的骤停。不幸的是,由于团队运输,安装或激活除颤设备的速度可能很慢,因此可能会出现过多的延迟。除颤延迟应被视为是真正的错误-医疗错误的危险来源。凭直觉,人们期望将大多数除颤延迟归因于人为错误。但是,现在人们越来越多地认识到有缺陷的设备设计对医疗系统错误的巨大影响。3我们试图通过以下系列证明有缺陷的除颤电极设计会直接导致难以忍受的除颤延迟。案例报告我们的机构使用Quik-Combo? (Medtronic,Redmond WA)除颤器电极或Padpro?除颤器电极(型号2516 Padpro LLC,美国密歇根州安阿伯)已连接到Lifepak? 12除颤器(Medtronic,华盛顿州雷德蒙德)。与Quik-Combo不同,Padpro电极需要定制的插头适配器才能串联在它和Lifepak 12治疗电缆之间(见图1)。我们确定医务人员对Padpro及其附加适配器的困惑造成了两名患者除颤的不合理延误。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号