Quality Control Tests in Some Diagnostic X-Ray Units in Bangladesh

机译:孟加拉国某些诊断X射线装置的质量控制测试

获取原文
           

摘要

Regular implementation of quality control in diagnostic x-ray facilities is essential to provide good quality images which lead to proper diagnoses with minimum hazard and distress. Important performance tests in diagnostic radiology in Bangladesh are carried out according to a quality control protocol and the measured parameter values are compared to the relevant acceptance limits. In this work, beam alignment, field congruence, nominal focal spot, film-screen contact and half value layer for diagnostic x-ray facilities are measured by using beam alignment test tool, RMI/Victoreen collimator test tool, Bar pattern focal spot test tool, film/screen contact test tool (RMI143D), Gammex RMI step wedge and densitometer from forty different diagnostic x-ray facilities in Bangladesh. For congruence between optical and radiation fields, 77.5% are found to be within limit and 60% of facilities are within the beam alignment limit. For most of the installations, 92.5% nominal focal spot size of diagnostic x-ray machines is matched perfectly with the rating of focal spot size. In an effort to improve image quality, this study has checked the film-screen contacts of multiple facilities and found 65% to have the expected uniformity. While investigating half value layers (HVL), a measure of x-ray beam quality, it is found that none of the diagnostic x-ray installations can achieve the recommended levels. DOI: http://dx.doi.org/10.3329/bjmp.v4i1.14688 Bangladesh Journal of Medical Physics Vol.4 No.1 2011 59-66
机译:定期在诊断X射线设备中实施质量控制对于提供高质量的图像至关重要,从而可以以最小的危害和困扰进行正确的诊断。孟加拉国诊断放射学中的重要性能测试是根据质量控制协议进行的,并将测得的参数值与相关的接受极限进行比较。在这项工作中,通过使用光束对准测试工具,RMI / Victoreen准直仪测试工具,条形图案焦点测试工具来测量用于诊断X射线设备的光束对准,场一致性,标称焦点,胶片屏幕接触和半值层,胶片/屏幕接触测试工具(RMI143D),来自孟加拉国40个不同X射线诊断设备的Gammex RMI阶梯楔形仪和光密度计。对于光场和辐射场之间的一致性,发现77.5%在限制之内,而60%的设施在光束对准限制之内。对于大多数安装,诊断X射线机的92.5%标称焦斑尺寸与焦斑尺寸额定值完美匹配。为了提高图像质量,本研究检查了多个设备的胶片屏幕接触,发现65%的胶片具有预期的均匀性。在研究半值层(HVL)(一种X射线束质量的量度)时,发现没有任何诊断X射线装置可以达到推荐的水平。 DOI:http://dx.doi.org/10.3329/bjmp.v4i1.14688孟加拉国医学物理学杂志2011年第4卷第1期59-66

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号