The effect of dorsal rim loss on the initial stability of the BioMedtrix cementless acetabular cup

机译:背缘缺失对BioMedtrix非骨水泥髋臼杯初始稳定性的影响

获取原文
           

摘要

Background Loss of dorsal acetabular rim (DAR) is a common sequela to canine hip dysplasia. The purpose of this study is to evaluate the effect of DAR loss on the initial stability of the cementless (BFX) acetabular cup. BFX cups were implanted into foam blocks reamed to resemble acetabulae with simulated 0, 25, 50, and 75% DAR loss. Models were tested in edge loading of the lateral surface of the cup with an indenter, and in centered loading with an articulated femoral prosthesis. Additionally, cups were implanted into paired cadaveric canine hemipelves with either no DAR depletion, or removal of 50% of the DAR, and acutely loaded to failure with an articulated femoral prosthesis. Results Mean load measured at 1?mm cup displacement during edge loading was not significantly different in foam blocks with loss of 0, 25, 50, and 75% DAR (360?±?124?N, 352?±?42?N, 330?±?81?N, 288?±?43?N, respectively; P?=?0.425). Mean load to failure with centered loads was greatest in blocks with 0% DAR loss (2828?±?208?N; P?Conclusions BFX cup stability under normal physiologic loads does not appear to be compromised in acetabulae with up to 50% DAR loss.
机译:背景髋臼背缘缺失(DAR)是犬髋关节发育不良的常见后遗症。这项研究的目的是评估DAR损失对非骨水泥(BFX)髋臼杯初始稳定性的影响。将BFX杯植入扩孔的泡沫块中,使其类似于髋臼,模拟的DAR损失为0、25、50和75%。使用压头在杯子侧面的边缘载荷和铰接式股骨假体的中心载荷下测试了模型。另外,将杯植入成对的尸体犬半骨中,没有DAR消耗,也没有切除50%的DAR,并通过关节铰接的股骨假体迅速加载至失败。结果泡沫块中,在边缘加载过程中杯位移为1?mm时测得的平均负载无明显差异,DAR损失为0、25、50和75%(360?±?124?N,352?±?42?N,分别为330±±81°N,288±±43°N;P≥0.425。 DAR损失为0%(2828?±?208?N; P?)时,以中心负荷为中心的平均失败至最大载荷结论BFX杯在正常生理负荷下的稳定性在DAR损失高达50%的髋臼中似乎没有受到损害。 。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号